We are developing a bolometer based on a differential thermometer that senses temperature changes through changes in the kinetic inductance of a superconducting thin film. The temperature transducer is an inductance bridge patterned as an integrated circuit on a 1 cm{sup 2} Si substrate. Two inductors from opposite arms of the bridge are patterned on a 2 mm{sup 2} thermally isolated Si island which is supported by a 9 {mu}m thick Si:B membrane. The bridge is excited with audio frequency current and the bridge imbalance is detected with a commercial DC SQUID amplifier. The bridge is balanced by applying power to the thermally isolated island. This thermometer is the sensor for a prototype radiometer that will provide an absolute measure of IR power.

OSTI ID:
5923530
Report Number(s):
CONF-900944-; CODEN: IEMGA
Journal Information:
IEEE Transactions on Magnetics (Institute of Electrical and Electronics Engineers); (United States), Vol. 27:2; Conference: 1990 applied superconductivity conference, Snowmass, CO (United States), 24-28 Sep 1990; ISSN 0018-9464