The energy calibration of x-ray absorption spectra using multiple-beam diffraction
- Fachhochschule Ostfriesland, Constantiaplatz 4, D-2970 Emden (Germany)
- Hamburger Synchrotronstrahlungslabor HASYLAB at DESY, Notkestrasse 85, D-2000 Hamburg 52 (Germany)
A new method for calibrating the energy scale of x-ray absorption spectra from an energy dispersive spectrometer has been developed. Distinct features in the diffracted intensity of the curved silicon crystal monochromator have been assigned to multiple-beam diffraction. The photon energies of these structures can be calculated if the precise spacing of the diffracting planes and the orientation of the crystal relative to the incident synchrotron radiation are known. The evaluation of Miller indices of operative reflections and the calculation of the corresponding photon energy is presented. The assignment of operative reflexes is simplified if the monochromator crystal can be rotated around the main diffracting vector {bold H}.
- OSTI ID:
- 5410561
- Journal Information:
- Review of Scientific Instruments; (United States), Vol. 63:1; ISSN 0034-6748
- Country of Publication:
- United States
- Language:
- English
Similar Records
Use of simultaneous reflections for precise absolute energy calibration of x rays
Measurement of synchrotron x-ray energies and line shapes using diffraction markers
Related Subjects
X-RAY SPECTROMETERS
CALIBRATION
ABSORPTION SPECTRA
DESY
MONOCHROMATORS
SYNCHROTRON RADIATION
YTTRIUM OXIDES
ACCELERATORS
BREMSSTRAHLUNG
CHALCOGENIDES
CYCLIC ACCELERATORS
ELECTROMAGNETIC RADIATION
MEASURING INSTRUMENTS
OXIDES
OXYGEN COMPOUNDS
RADIATIONS
SPECTRA
SPECTROMETERS
SYNCHROTRONS
TRANSITION ELEMENT COMPOUNDS
YTTRIUM COMPOUNDS
440101* - Radiation Instrumentation- General Detectors or Monitors & Radiometric Instruments