Title: Time-resolved experiments in the frequency domain using synchrotron radiation (invited)

Journal Article · · Review of Scientific Instruments; (United States)
DOI:https://doi.org/10.1063/1.1143025· OSTI ID:5401009
 [1]; ; ;  [2];  [3]
  1. Istituto di Struttura della Materia del Consiglio Nazionale delle Ricerche, Via Enrico Fermi, 38, 00044, Frascati, Rome (Italy)
  2. Istituto di Medicina Sperimentale del Consiglio Nazionale delle Ricerche, Viale Marx, 15, 00137, Rome (Italy)
  3. Laboratorio di Tossicologia Comparata ed Ecotossicologia, Istituto Superiore di Sanita', Viale Regina Elena, 279, 00161, Rome (Italy)

PLASTIQUE is the only synchrotron radiation beam line in the world that performs time-resolved fluorescence experiments in frequency domain. These experiments are extremely valuable sources of information on the structure and the dynamics of molecules. This technique measures fluorescence lifetimes with picosecond resolution in the near UV spectral range. Such accurate measurements are rendered possible by taking phase and modulation data, and by the advantages of the cross-correlation technique. A successful experiment demonstrated the radiation damage induced by low doses of radiation on rabbit blood cell membranes.

OSTI ID:
5401009
Journal Information:
Review of Scientific Instruments; (United States), Vol. 63:1; ISSN 0034-6748
Country of Publication:
United States
Language:
English