skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Accuracy of measurements of the quantized Hall resistivity by a direct current comparator type potentiometer; Calibration using a Josephson potentiometer

Journal Article · · IEEE (Institute of Electrical and Electronics Engineers) Transactions on Instrumentation and Measurement; (USA)
DOI:https://doi.org/10.1109/19.192287· OSTI ID:5147527
; ;  [1]
  1. Electrotechnical Lab., 1-1-4 Umezono, Tsukuba-shi, Ibaraki 305 (JP)

In this paper the accuracy of measurements of the quantized Hall resistivity (QHR) by a direct current comparator type (DCC) potentiometer is estimated based on a linearity calibration of the potentiometer using a Josephson potentiometer. The power coefficient contribution to the nonlinearity of the DCC potentiometer is found to be 0.15 +- 0.02 ppm/(100 mV)/sup 2/ at 18.5 +-0.5{sup 0}C in 1985 and 0.21 +- 0.03 ppm/ (100 mV)/sup 2/ at 20.5 +-0.5{sup 0}C in 1988. The possibility of accurate measurements of the ratio QHR/R/sub std/ with uncertainties less than 0.05 ppm by the DCC potentiometer is discussed.

OSTI ID:
5147527
Journal Information:
IEEE (Institute of Electrical and Electronics Engineers) Transactions on Instrumentation and Measurement; (USA), Vol. 38:2; ISSN 0018-9456
Country of Publication:
United States
Language:
English