Title: Testing the performance of real-time incinerator emission monitors. Project report

Technical Report ·
OSTI ID:464284

Ten prototypes of continuous emission monitors (CEMs) for measuring trace metal or trace organic species concentrations were tested. Of the 10 CEMs tested, four measured incinerator flue gas concentrations of several specific volatile organic compounds (VOCs), one measured total particulate-bound polynuclear aromatic hydrocarbon (PAH) concentrations, two measured flue gas concentrations of several (up to 14) trace metals, and three measured mercury concentrations. While the testing consisted of obtaining quantitative measurement data on the four measures of CEM performance checked in a relative accuracy test audit (RATA) as described in 40 CFR 60 Appendix F -- relative accuracy (RA), calibration drift (CD), zero drift (ZD), and response time - the primary project objective focused on the RA measurement. Four series of tests were performed, each simultaneously testing up to three monitors measuring the same or similar analyte type. Each test series consisted of performing triplicate Reference Method (RM) measurements at each of three target flue gas monitored analyte concentrations while the tested CEMs were in operation.

Research Organization:
Acurex Corp., Jefferson, AR (United States). Incineration Reserach Facility
OSTI ID:
464284
Report Number(s):
PB-97-142871/XAB; CNN: Contract EPA-68-C4-0044; TRN: 70911631
Resource Relation:
Other Information: PBD: Mar 1997