Title: Dose conversion factors and linear energy transfer for irradiation of thin blood layers with low-energy X rays

Journal Article · · Radiation Research
DOI:https://doi.org/10.2307/3578785· OSTI ID:45261

For irradiation of thin samples of biological material with low-energy X rays, conversion of measured air kerma, free in air to average absorbed dose to the sample is necessary. In the present paper, conversion factors from measured air kerma to average absorbed dose in thin blood samples are given for four low-energy X-ray qualities (14-50 kVp). These factors were obtained by Monte Carlo simulation of a practical sample holder. Data for different thicknesses of the blood and backing layer are presented. The conversion factors are found to depend strongly on the thicknesses of the blood layer and backing layer. In radiobiological work, knowledge of linear energy transfer (LET) values for the radiation quality used is often required. Track-averaged LET values for low-energy X rays are presented in this work. It is concluded that the thickness of the sample does not influence the LET value appreciably, indicating that for all radiobiological purposes this value can be regarded as a constant throughout the sample. Furthermore, the large difference between the LET value for a 50 kV spectrum found in this work and the value given in ICRU Report 16 is pointed out. 16 refs., 7 figs., 1 tab.

Sponsoring Organization:
USDOE
OSTI ID:
45261
Journal Information:
Radiation Research, Vol. 137, Issue 1; Other Information: PBD: Jan 1994
Country of Publication:
United States
Language: