PVDF piezofilm transducer for measuring the parameters of shock waves in focused systems
A description is given of the construction and testing of a pressure transducer for measuring shock wave parameters in water, as applied to lithotripters and other focused systems. The sensitive element is made from a PVDF piezofilm, with a working area of 2 mm diameter. The transducer sensitivity is 0.2 V/MPa, and the rise time is not greater than 50 nsec.
- Sponsoring Organization:
- USDOE
- OSTI ID:
- 263355
- Journal Information:
- Measurement Techniques, Vol. 38, Issue 6; Other Information: PBD: Nov 1995; TN: Translated from Izmeritel`naya Tekhnika; No. 6, 58-59(Jun 1995)
- Country of Publication:
- United States
- Language:
- English