Title: SU-E-I-40: Phantom Research On Monochromatic Images Taken by Dual CBCT with Multiple Energy Sets

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4924037· OSTI ID:22493997
 [1];  [2];  [1];  [3]; ;  [4]
  1. Duke University, Durham, NC - North Carolina (United States)
  2. Shandong University, Jinan, Shandong (China)
  3. The Affiliated Cancer Hospital of Zhengzhou University, Zhengzhou, Henan (China)
  4. Duke University Medical Center, Durham, NC (United States)

Purpose: To evaluate the quality of monochromatic images at the same virtual monochromatic energy using dual cone-beam computed tomography (CBCT) with either kV/kV or MV/kV or MV/MV energy sets. Methods: CT images of Catphan 504 phantom were acquired using four different KV and MV settings: 80kV, 140kV, 4MV, 6MV. Three sets of monochromatic images were calculated: 80kV-140kV, 140kV-4MV and 4MV-6MV. Each set of CBCT images were reconstructed from the same selected virtual monochromatic energy of 1MeV. Contrast-to-Noise Ratios (CNRs) were calculated and compared between each pair of images with different energy sets. Results: Between kV/MV and MV/MV images, the CNRs are comparable for all inserts. However, differences of CNRs were observed between the kV/kV and kV/MV images. Delrin’s CNR ratio between kV/kV image and kV/MV image is 1.634. LDPE’s (Low-Density Polyethylene) CNR ratio between kV/kV and kV/MV images is 0.509. Polystyrene’s CNR ratio between kV/kV image and kV/MV image is 2.219. Conclusion: Preliminary results indicated that the CNRs calculated from CBCT images reconstructed from either kV/MV projections or MV/MV projections for the same selected virtual monochromatic energy may be comparable.

OSTI ID:
22493997
Journal Information:
Medical Physics, Vol. 42, Issue 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English

Similar Records

Generation of virtual monochromatic CBCT from dual kV/MV beam projections
Journal Article · Sun Dec 15 00:00:00 EST 2013 · Medical Physics · OSTI ID:22493997

Generation of virtual monochromatic CBCT from dual kV/MV beam projections
Journal Article · Sun Dec 15 00:00:00 EST 2013 · Medical Physics · OSTI ID:22493997

WE-AB-207A-07: A Planning CT-Guided Scatter Artifact Correction Method for CBCT Images
Journal Article · Wed Jun 15 00:00:00 EDT 2016 · Medical Physics · OSTI ID:22493997