skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: SU-E-I-24: Method for CT Automatic Exposure Control Verification

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4924021· OSTI ID:22493983

Purpose: Design of a phantom and a simple method for the automatic exposure control (AEC) verification in CT. This verification is included in the computed tomography (CT) Spanish Quality Assurance Protocol. Methods: The phantom design is made from the head and the body phantom used for the CTDI measurement and PMMA plates (35×35 cm2) of 10 cm thickness. Thereby, three different thicknesses along the longitudinal axis are obtained which permit to evaluate the longitudinal AEC performance. Otherwise, the existent asymmetry in the PMMA layers helps to assess angular and 3D AEC operation.Recent acquisition in our hospital (August 2014) of Nomex electrometer (PTW), together with the 10 cm pencil ionization chamber, led to register dose rate as a function of time. Measurements with this chamber fixed at 0° and 90° on the gantry where made on five multidetector-CTs from principal manufacturers. Results: Individual analysis of measurements shows dose rate variation as a function of phantom thickness. The comparative analysis shows that dose rate is kept constant in the head and neck phantom while the PMMA phantom exhibits an abrupt variation between both results, being greater results at 90° as the thickness of the phantom is 3.5 times larger than in the perpendicular direction. Conclusion: Proposed method is simple, quick and reproducible. Results obtained let a qualitative evaluation of the AEC and they are consistent with the expected behavior. A line of future development is to quantitatively study the intensity modulation and parameters of image quality, and a possible comparative study between different manufacturers.

OSTI ID:
22493983
Journal Information:
Medical Physics, Vol. 42, Issue 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English