Title: SU-D-210-07: The Dependence On Acoustic Velocity of Medium On the Needle Template and Electronic Grid Alignment in Ultrasound QA for Prostate Brachytherapy

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4923922· OSTI ID:22486647
; ;  [1]
  1. Virginia Commonwealth University, Richmond, VA (United States)

Purpose: To analyze the impact on acoustic velocity (AV) of two different media (water and milk) using the needle template/electronic grid alignment test. Water, easily available, makes a good material to test the alignment of the template and grid although water’s AV (1498 m/s at 25°C) is significantly different from tissue (1540 m/s). Milk, with an AV much closer (1548 m/s) to prostate tissue, may be a good substitute for water in ultrasound quality assurance testing. Methods: Tests were performed using a Hitachi ultrasound unit with a mechanical arrangement designed to position needles parallel to the transducer. In this work, two materials – distilled water and homogenized whole milk (AVs of 1498 and 1548 m/s at 25°C) were used in a phantom to test ultrasound needle/grid alignment. The images were obtained with both materials and analyzed for their placement accuracy. Results: The needle template/electronic grid alignment tests showed displacement errors between measured and calculated values. The measurements showed displacements of 2.3mm (water) and 0.4mm (milk), and 1.6mm (water) and 0.3mm (milk) at depths of 7cm and 5cm respectively from true needle positions. The calculated results showed a displacement of 2.36 mm (water); 0.435mm (milk), and 1.66mm (water) and 0.31mm (milk) at a depth of 7cm and 5cm respectively. The displacements in the X and Y directions were also calculated. At depths of 7cm and 5cm, the (ΔX,ΔY) displacements in water were (0.829mm, 2.21mm) and (0.273mm, 1.634mm) and for milk were (0.15mm, 0.44mm) and (0.05mm, 0.302mm) respectively. Conclusion: The measured and calculated values were in good agreement for all tests. They show that milk provides superior results when performing needle template and electronic grid alignment tests for ultrasound units used in prostate brachytherapy.

OSTI ID:
22486647
Journal Information:
Medical Physics, Vol. 42, Issue 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English