Title: SU-D-BRD-02: Auto Weekly - An Automated Online Weekly Chart Check System for Medical Physics

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4923868· OSTI ID:22486609

Purpose: Radiation therapy has evolved to become not only more precise and potent, but also more complicated to monitor and deliver. More rigorous and comprehensive quality assurance is needed to safeguard ever advancing radiation therapy. ICRU standards dictate that an ever growing set of treatment parameters are manually checked weekly by medical physicists. This “weekly chart check” procedure is laborious and subject to human errors or other factors. A computer-assisted chart checking process will enable more complete and accurate human review of critical parameters, reduce the risk of medical errors, and improve the efficiency. Methods: We developed a web-based software system that enables a thorough weekly quality assurance checks. In the backend, the software retrieves all machine parameters from a Treatment Management System (TMS) and compares them against the corresponding ones from the treatment planning system. They are also checked for validity against preset rules. The results are displayed as a web page in the front-end for physicists to review. Then a summary report is generated and uploaded automatically to the TMS as a record for weekly chart checking. Results: The software system has been deployed on a web server in our department’s intranet, and has been tested thoroughly by our clinical physicists. A plan parameter would be highlighted when it is off the preset limit. The developed system has changed the way of checking charts with significantly improved accuracy, efficiency, and completeness. It has been shown to be robust, fast, and easy to use. Conclusion: A computer-assisted system has been developed for efficient, accurate, and comprehensive weekly chart checking. The system has been extensively validated and is being implemented for routine clinical use.

OSTI ID:
22486609
Journal Information:
Medical Physics, Vol. 42, Issue 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English

Similar Records

SU-D-BRD-01: An Automated Physics Weekly Chart Checking System Supporting ARIA
Journal Article · Mon Jun 15 00:00:00 EDT 2015 · Medical Physics · OSTI ID:22486609

A computer aided treatment event recognition system in radiation therapy
Journal Article · Wed Jan 15 00:00:00 EST 2014 · Medical Physics · OSTI ID:22486609