Title: Multihelix rotating shield brachytherapy for cervical cancer

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4933244· OSTI ID:22482402
 [1]; ;  [2]
  1. Department of Biomedical Engineering, University of Iowa, 1402 Seamans Center for the Engineering Arts and Sciences, Iowa City, Iowa 52242 (United States)
  2. Department of Radiation Oncology, University of Iowa, 200 Hawkins Drive, Iowa City, Iowa 52242 and Department of Electrical and Computer Engineering, University of Iowa, 4016 Seamans Center for the Engineering Arts and Sciences, Iowa City, Iowa 52242 (United States)

Purpose: To present a novel brachytherapy technique, called multihelix rotating shield brachytherapy (H-RSBT), for the precise angular and linear positioning of a partial shield in a curved applicator. H-RSBT mechanically enables the dose delivery using only linear translational motion of the radiation source/shield combination. The previously proposed approach of serial rotating shield brachytherapy (S-RSBT), in which the partial shield is rotated to several angular positions at each source dwell position [W. Yang et al., “Rotating-shield brachytherapy for cervical cancer,” Phys. Med. Biol. 58, 3931–3941 (2013)], is mechanically challenging to implement in a curved applicator, and H-RSBT is proposed as a feasible solution. Methods: A Henschke-type applicator, designed for an electronic brachytherapy source (Xoft Axxent™) and a 0.5 mm thick tungsten partial shield with 180° or 45° azimuthal emission angles and 116° asymmetric zenith angle, is proposed. The interior wall of the applicator contains six evenly spaced helical keyways that rigidly define the emission direction of the partial radiation shield as a function of depth in the applicator. The shield contains three uniformly distributed protruding keys on its exterior wall and is attached to the source such that it rotates freely, thus longitudinal translational motion of the source is transferred to rotational motion of the shield. S-RSBT and H-RSBT treatment plans with 180° and 45° azimuthal emission angles were generated for five cervical cancer patients with a diverse range of high-risk target volume (HR-CTV) shapes and applicator positions. For each patient, the total number of emission angles was held nearly constant for S-RSBT and H-RSBT by using dwell positions separated by 5 and 1.7 mm, respectively, and emission directions separated by 22.5° and 60°, respectively. Treatment delivery time and tumor coverage (D{sub 90} of HR-CTV) were the two metrics used as the basis for evaluation and comparison. For all the generated treatment plans, the D{sub 90} of the HR-CTV in units of equivalent dose in 2 Gy fractions (EQD2) was escalated until the D{sub 2cc} (minimum dose to hottest 2 cm{sup 3}) tolerance of either the bladder (90 Gy{sub 3}), rectum (75 Gy{sub 3}), or sigmoid colon (75 Gy{sub 3}) was reached. Results: Treatment time changed for H-RSBT versus S-RSBT by −7.62% to 1.17% with an average change of −2.8%, thus H-RSBT treatments times tended to be shorter than for S-RSBT. The HR-CTV D{sub 90} also changed by −2.7% to 2.38% with an average of −0.65%. Conclusions: H-RSBT is a mechanically feasible delivery technique for use in the curved applicators needed for cervical cancer brachytherapy. S-RSBT and H-RSBT were clinically equivalent for all patients considered, with the H-RSBT technique tending to require less time for delivery.

OSTI ID:
22482402
Journal Information:
Medical Physics, Vol. 42, Issue 11; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English

Similar Records

SU-E-T-564: Multi-Helix Rotating Shield Brachytherapy for Cervical Cancer
Journal Article · Mon Jun 15 00:00:00 EDT 2015 · Medical Physics · OSTI ID:22482402

Dynamic rotating-shield brachytherapy
Journal Article · Sun Dec 15 00:00:00 EST 2013 · Medical Physics · OSTI ID:22482402

Paddle-based rotating-shield brachytherapy
Journal Article · Thu Oct 15 00:00:00 EDT 2015 · Medical Physics · OSTI ID:22482402