Title: Is high–dose rate RapidArc-based radiosurgery dosimetrically advantageous for the treatment of intracranial tumors?

Journal Article · · Medical Dosimetry

In linac-based stereotactic radiosurgery (SRS) and radiotherapy (SRT), circular cone(s) or conformal arc(s) are conventionally used to treat intracranial lesions. However, when the target is in close proximity to critical structures, it is frequently quite challenging to generate a quality plan using these techniques. In this study, we investigated the dosimetric characteristics of using high–dose rate RapidArc (RA) technique for radiosurgical treatment of intracranial lesions. A total of 10 intracranial SRS/SRT cases previously planned using dynamic conformal arc (DCA) or cone-based techniques have been included in this study. For each case, 3 treatment plans were generated: (1) a DCA plan with multiple noncoplanar arcs, (2) a high–dose rate RA plan with arcs oriented the same as DCA (multiple-arc RA), and 3) a high–dose rate RA plan with a single coplanar arc (single-arc RA). All treatment plans were generated under the same prescription and similar critical structure dose limits. Plan quality for different plans was evaluated by comparing various dosimetric parameters such as target coverage, conformity index (CI), homogeneity index (HI), critical structures, and normal brain tissue doses as well as beam delivery time. With similar critical structure sparing, high–dose rate RA plans can achieve much better target coverage, dose conformity, and dose homogeneity than the DCA plans can. Plan quality indices CI and HI, for the DCA, multiple-arc RA, and single-arc RA techniques, were measured as 1.67 ± 0.39, 1.32 ± 0.28, and 1.38 ± 0.30 and 1.24 ± 0.11, 1.10 ± 0.04, and 1.12 ± 0.07, respectively. Normal brain tissue dose (V{sub 12} {sub Gy}) was found to be similar for DCA and multiple-arc RA plans but much larger for the single-arc RA plans. Beam delivery was similar for DCA and multiple-arc RA plans but shorter with single-arc RA plans. Multiple-arc RA SRS/SRT can provide better treatment plans than conventional DCA plans, especially for complex cases.

OSTI ID:
22462416
Journal Information:
Medical Dosimetry, Vol. 40, Issue 1; Other Information: Copyright (c) 2015 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.; Country of input: International Atomic Energy Agency (IAEA); ISSN 0958-3947
Country of Publication:
United States
Language:
English