Development of a wide-range tritium-concentration detector
- China Academy of Engineering Physics, Mianyang (China)
According to the requirements of the tritium related systems of the TBM (Test Blanket Module) for monitoring the on-line tritium concentration, a wide-range tritium-concentration detector has been developed to measure the tritium concentration in the range of 10{sup 4} Bq/ml - 5*10{sup 8} Bq/ml. This detector is combined with a low-memory helium ionization chamber. The weak current signal collected in the ionization chamber is converted to the voltage signal by an I-V converter. The minimum weak current which the detector could be measured is 10{sup -14} A. The performance of the background current and the current response linearity of the prototype have been tested. The test result indicates that the linear response of the current signal of the prototype without connecting the ionization chamber is good. The linear correlation coefficient is R{sup 2} = 0.998.
- OSTI ID:
- 22429711
- Journal Information:
- Fusion Science and Technology, Vol. 67, Issue 2; Conference: TRITIUM 2013: 10. International Conference on Tritium Science and Technology, Nice Acropolis (France), 21-25 Oct 2013; Other Information: Country of input: France; 3 refs.; ISSN 1536-1055
- Country of Publication:
- United States
- Language:
- English
Similar Records
Wide-range tritium monitor
Electret ion chambers for passive measurement of airborne tritium: Theory and practice