Development and Clinical Implementation of a Universal Bolus to Maintain Spot Size During Delivery of Base of Skull Pencil Beam Scanning Proton Therapy
- Department of Radiation Oncology, University of Pennsylvania, Philadelphia, Pennsylvania (United States)
Purpose: To report on a universal bolus (UB) designed to replace the range shifter (RS); the UB allows the treatment of shallow tumors while keeping the pencil beam scanning (PBS) spot size small. Methods and Materials: Ten patients with brain cancers treated from 2010 to 2011 were planned using the PBS technique with bolus and the RS. In-air spot sizes of the pencil beam were measured and compared for 4 conditions (open field, with RS, and with UB at 2- and 8-cm air gap) in isocentric geometry. The UB was applied in our clinic to treat brain tumors, and the plans with UB were compared with the plans with RS. Results: A UB of 5.5 cm water equivalent thickness was found to meet the needs of the majority of patients. By using the UB, the PBS spot sizes are similar with the open beam (P>.1). The heterogeneity index was found to be approximately 10% lower for the UB plans than for the RS plans. The coverage for plans with UB is more conformal than for plans with RS; the largest increase in sparing is usually for peripheral organs at risk. Conclusions: The integrity of the physical properties of the PBS beam can be maintained using a UB that allows for highly conformal PBS treatment design, even in a simple geometry of the fixed beam line when noncoplanar beams are used.
- OSTI ID:
- 22420411
- Journal Information:
- International Journal of Radiation Oncology, Biology and Physics, Vol. 90, Issue 1; Other Information: Copyright (c) 2014 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.; Country of input: International Atomic Energy Agency (IAEA); ISSN 0360-3016
- Country of Publication:
- United States
- Language:
- English
Similar Records
SU-E-T-375: Passive Scattering to Pencil-Beam-Scanning Comparison for Medulloblastoma Proton Therapy: LET Distributions and Radiobiological Implications
SU-E-T-441: Comparison of Dose Distributions for Spot-Scanned Pencil-Beam and Scattered-Beam Proton Treatments of Ocular Tumors