Abstract

Purpose: Knowing the contribution of neutron to collateral effects in treatments is both a complex and a mandatory task. This work aims to present an operative procedure for neutron estimates in any facility using a neutron digital detector. Methods: The authors’ previous work established a linear relationship between the total second cancer risk due to neutrons (TR{sup n}) and the number of MU of the treatment. Given that the digital detector also presents linearity with MU, its response can be used to determine the TR{sup n} per unit MU, denoted as m, normally associated to a generic Linac model and radiotherapy facility. Thus, from the number of MU of each patient treatment, the associated risk can be estimated. The feasibility of the procedure was tested by applying it in eight facilities; patients were evaluated as well. Results: From the reading of the detector under selected irradiation conditions, m values were obtained for different machines, ranging from 0.25 × 10{sup −4}% per MU for an Elekta Axesse at 10 MV to 6.5 × 10{sup −4}% per MU for a Varian Clinac at 18 MV. Using these values, TR{sup n} of patients was estimated in each facility and compared to that frommore » the individual evaluation. Differences were within the range of uncertainty of the authors’ methodology of equivalent dose and risk estimations. Conclusions: The procedure presented here allows an easy estimation of the second cancer risk due to neutrons for any patient, given the number of MU of the treatment. It will enable the consideration of this information when selecting the optimal treatment for a patient by its implementation in the treatment planning system.« less

Authors:
 [1];  [2];  [3];  [4];  [5];  [6];  [7];  [8];  [9];
  1. Instituto de Física, Pontificia Universidad Católica de Chile, Santiago 4880 (Chile)
  2. Servicio de Radiofísica, Hospital Universitario Virgen Macarena, Sevilla 41009 (Spain)
  3. Serviço de Física Médica, Instituto Português de Oncologia, Coimbra 3000-075 (Portugal)
  4. i3N, Department of Physics, University of Aveiro, Aveiro 3810-193 (Portugal)
  5. Radiotherapy Service, Russian Research Center for Radiology and Surgical Technology, Saint Petersburg 197758 (Russian Federation)
  6. Elekta, Ltd., Crawley RH10 9RR (United Kingdom)
  7. Servicio de Radiofísica, Instituto Onkologikoa, San Sebastián 20014 (Spain)
  8. Servicio de Radiofísica, Hospital Universitario de la Ribera, Alzira 46600, Valencia (Spain)
  9. Departament de Física, Universitat Autònoma de Barcelona, Bellaterra 08193 (Spain)
Publication Date:
OSTI Identifier:
22413381
Resource Type:
Journal Article
Journal Name:
Medical Physics
Additional Journal Information:
Journal Volume: 42; Journal Issue: 1; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); Journal ID: ISSN 0094-2405
Country of Publication:
United States
Language:
English
Subject:
60 APPLIED LIFE SCIENCES; COMMISSIONING; COMPARATIVE EVALUATIONS; HAZARDS; LINEAR ACCELERATORS; NEOPLASMS; NEUTRONS; PATIENTS; RADIOTHERAPY

Citation Formats

Romero-Expósito, M., E-mail: mariateresa.romero@uab.cat, Sánchez-Nieto, B., Terrón, J. A., Lopes, M. C., Ferreira, B. C., Grishchuk, D., Sandín, C., Moral-Sánchez, S., Melchor, M., Domingo, C., and others, and. Commissioning the neutron production of a Linac: Development of a simple tool for second cancer risk estimation. United States: N. p., 2015. Web. doi:10.1118/1.4903525.
Romero-Expósito, M., E-mail: mariateresa.romero@uab.cat, Sánchez-Nieto, B., Terrón, J. A., Lopes, M. C., Ferreira, B. C., Grishchuk, D., Sandín, C., Moral-Sánchez, S., Melchor, M., Domingo, C., & others, and. Commissioning the neutron production of a Linac: Development of a simple tool for second cancer risk estimation. United States. https://doi.org/10.1118/1.4903525
Romero-Expósito, M., E-mail: mariateresa.romero@uab.cat, Sánchez-Nieto, B., Terrón, J. A., Lopes, M. C., Ferreira, B. C., Grishchuk, D., Sandín, C., Moral-Sánchez, S., Melchor, M., Domingo, C., and others, and. 2015. "Commissioning the neutron production of a Linac: Development of a simple tool for second cancer risk estimation". United States. https://doi.org/10.1118/1.4903525.
@article{osti_22413381,
title = {Commissioning the neutron production of a Linac: Development of a simple tool for second cancer risk estimation},
author = {Romero-Expósito, M., E-mail: mariateresa.romero@uab.cat and Sánchez-Nieto, B. and Terrón, J. A. and Lopes, M. C. and Ferreira, B. C. and Grishchuk, D. and Sandín, C. and Moral-Sánchez, S. and Melchor, M. and Domingo, C. and others, and},
abstractNote = {Purpose: Knowing the contribution of neutron to collateral effects in treatments is both a complex and a mandatory task. This work aims to present an operative procedure for neutron estimates in any facility using a neutron digital detector. Methods: The authors’ previous work established a linear relationship between the total second cancer risk due to neutrons (TR{sup n}) and the number of MU of the treatment. Given that the digital detector also presents linearity with MU, its response can be used to determine the TR{sup n} per unit MU, denoted as m, normally associated to a generic Linac model and radiotherapy facility. Thus, from the number of MU of each patient treatment, the associated risk can be estimated. The feasibility of the procedure was tested by applying it in eight facilities; patients were evaluated as well. Results: From the reading of the detector under selected irradiation conditions, m values were obtained for different machines, ranging from 0.25 × 10{sup −4}% per MU for an Elekta Axesse at 10 MV to 6.5 × 10{sup −4}% per MU for a Varian Clinac at 18 MV. Using these values, TR{sup n} of patients was estimated in each facility and compared to that from the individual evaluation. Differences were within the range of uncertainty of the authors’ methodology of equivalent dose and risk estimations. Conclusions: The procedure presented here allows an easy estimation of the second cancer risk due to neutrons for any patient, given the number of MU of the treatment. It will enable the consideration of this information when selecting the optimal treatment for a patient by its implementation in the treatment planning system.},
doi = {10.1118/1.4903525},
url = {https://www.osti.gov/biblio/22413381}, journal = {Medical Physics},
issn = {0094-2405},
number = 1,
volume = 42,
place = {United States},
year = {Thu Jan 15 00:00:00 EST 2015},
month = {Thu Jan 15 00:00:00 EST 2015}
}