Title: SU-C-19A-01: A Simple Deep Inspiration Breath Hold System

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4889700· OSTI ID:22412426
; ; ;  [1]; ; ; ;  [2]
  1. Marquette General Hospital, Marquette, MI (United States)
  2. Dickinson County Healthcare System, Iron Mountain, MI (United States)

Purpose: Deep Inspiration Breath Hold (DIBH) Radiation therapy for left sided breast can reduce dose to the lungs and heart. The purpose of this work is to illustrate how to implement a simple method of DIBH for simulation and treatment using equipment readily available in most radiation oncology clinics. Methods: For simulation and treatment, a foam block is placed on the patient's abdomen or chest and a horizontal laser mounted on a movable slide is aimed at the center of the foam block. After a coaching session the block is marked at the average free breathing position and average DIBH position. The position of block relative to laser can be seen by the patient via prism glasses as well as the radiation therapists via a video camera system. Simulation CT scans and treatment delivery are performed under DIBH conditions. Imaging and treatment are performed by manually turning the beam on once the patient has achieved DIBH after being given verbal instructions. Results: Manually triggered imaging was used daily to verify DIBH reproducibility for all patients treated using this system. Sets of before and during port images were used to ensure patient position was appropriate for treatment. Results of the laser on block method were compared to a sister facility using surface mapping techniques for DIBH and the two methods were found to have clinically equivalent reproducibility. Conclusion: The laser and block system was found to be simple to implement and robust during patient treatment. This system can be created from readily available materials at low cost and provides adequate feedback to patient and therapists. During treatment images document the reproducibility of setup and give confidence to clinicians that this method is reproducible from day to day.

OSTI ID:
22412426
Journal Information:
Medical Physics, Vol. 41, Issue 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English