Title: WE-D-17A-03: Improvement of Accuracy of Spot-Scanning Proton Beam Delivery for Liver Tumor by Real-Time Tumor-Monitoring and Gating System: A Simulation Study

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4889406· OSTI ID:22407888
; ; ; ; ; ; ;  [1]; ;  [2];  [3]
  1. Graduate School of Medicine, Hokkaido University, Sapporo, Hokkaido (Japan)
  2. Hitachi, Ltd., Hitachi Research Laboratory, Hitachi, Ibaraki (Japan)
  3. Department of Engineering, Hokkaido University, Sapporo, Hokkaido (Japan)

Purpose: To improve the accuracy of spot-scanning proton beam delivery for target in motion, a real-time tumor-monitoring and gating system using fluoroscopy images was developed. This study investigates the efficacy of this method for treatment of liver tumors using simulation. Methods: Three-dimensional position of a fiducial marker inserted close to the tumor is calculated in real time and proton beam is gated according to the marker's distance from the planned position (Shirato, 2012). The efficient beam delivery is realized even for the irregular and sporadic motion signals, by employing the multiple-gated irradiations per operation cycle (Umezawa, 2012). For each of two breath-hold CTs (CTV=14.6cc, 63.1cc), dose distributions were calculated with internal margins corresponding to freebreathing (FB) and real-time gating (RG) with a 2-mm gating window. We applied 8 trajectories of liver tumor recorded during the treatment of RTRT in X-ray therapy and 6 initial timings. Dmax/Dmin in CTV, mean liver dose (MLD), and irradiation time to administer 3 Gy (RBE) dose were estimated assuming rigid motion of targets by using in-house simulation tools and VQA treatment planning system (Hitachi, Ltd., Tokyo). Results: Dmax/Dmin was degraded by less than 5% compared to the prescribed dose with all motion parameters for smaller CTV and less than 7% for larger CTV with one exception. Irradiation time showed only a modest increase if RG was used instead of FB; the average value over motion parameters was 113 (FB) and 138 s (RG) for smaller CTV and 120 (FB) and 207 s (RG) for larger CTV. In RG, it was within 5 min for all but one trajectory. MLD was markedly decreased by 14% and 5–6% for smaller and larger CTVs respectively, if RG was applied. Conclusions: Spot-scanning proton beam was shown to be delivered successfully to liver tumor without much lengthening of treatment time. This research was supported by the Cabinet Office, Government of Japan and the Japan Society for the Promotion of Science (JSPS) through the Funding Program for World-Leading Innovative R and D on Science and Technology (FIRST Program), initiated by the Council for Science and Technology Policy (CSTP)

OSTI ID:
22407888
Journal Information:
Medical Physics, Vol. 41, Issue 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405