Title: Precise and direct method for the measurement of the torsion spring constant of the atomic force microscopy cantilevers

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.4904866· OSTI ID:22392315
 [1]
  1. Institute of Fundamental Technological Research, Pawińskiego 5b, 02-106 Warsaw (Poland)

A direct method for the evaluation of the torsional spring constants of the atomic force microscope cantilevers is presented in this paper. The method uses a nanoindenter to apply forces at the long axis of the cantilever and in the certain distance from it. The torque vs torsion relation is then evaluated by the comparison of the results of the indentations experiments at different positions on the cantilever. Next, this relation is used for the precise determination of the torsional spring constant of the cantilever. The statistical analysis shows that the standard deviation of the calibration measurements is equal to approximately 1%. Furthermore, a simple method for calibration of the photodetector’s lateral response is proposed. The overall procedure of the lateral calibration constant determination has the accuracy approximately equal to 10%.

OSTI ID:
22392315
Journal Information:
Review of Scientific Instruments, Vol. 86, Issue 1; Other Information: (c) 2014 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748