skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: SU-E-T-617: A Feasibility Study of Navigation Based Multi Criteria Optimization for Advanced Cervical Cancer IMRT Planning

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4888953· OSTI ID:22369732
 [1]
  1. Shandong Cancer Hospital, Jinan, Shandong (China)

Purpose: This study aims to validate multi-criteria optimization (MCO) against standard intensity modulated radiation therapy (IMRT) optimization for advanced cervical cancer in RayStation (v2.4, RaySearch Laboratories, Sweden). Methods: 10 advanced cervical cancer patients IMRT plans were randomly selected, these plans were designed with step and shoot optimization, new plans were then designed with MCO based on these plans,while keeping optimization conditions unchanged,comparison was made between both kinds of plans including the dose volume histogram parameters of PTV and OAR,and were analysed by pairing-t test. Results: We normalize the plan so that 95% volume of PTV achieved the prescribed dose(50Gy). The volume of radiation 10, 20, 30, and 40 Gy of the rectum were reduced by 14.7%,26.8%,21.1%,10.5% respectively(P≥0.05). The mean dose of rectum were reduced by 7.2Gy(P≤0.05). There were no significant differences for the dosimetric parameters for the bladder. Conclusion: In comparision with standard IMRT optimization, MCO reduces the dose of organs at risk with the same PTV coverage,but the result needs further clinical evalution.

OSTI ID:
22369732
Journal Information:
Medical Physics, Vol. 41, Issue 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English