skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: SU-E-T-545: A MLC-Equipped Robotic Radiosurgery-Radiotherapy Combined System in Treating Hepatic Lesions: Delivery Efficiency as Compared to a Standard Linac for Treating Hepatic Lesions

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4888879· OSTI ID:22369666
; ; ; ;  [1];  [2]
  1. Fox Chase Cancer Center, Philadephia, PA (United States)
  2. Virtua Fox Chase Cancer Center, Philadelphia, PA (United States)

Purpose: The CyberKnife (CK) M6 Series introduced a mulitleaf collimator (MLC) beam for extending its capability to the conventional radiotherapy. This work is to investigate delivery efficiency of this system as compared to a standard Varian linac when treating hepatic lesions. Methods: Nine previously treated patients were divided into three groups with three patients in each. Group one: fractionated radiotherapy; Group two: SBRT-like treatments and Group three: fractionated radiotherapy targeting two PTVs. The clinically used plans were generated with the Eclipse treatment planning system (TPS). We re-planned these cases using a Mulitplan (MP) TPS for the CK M6 and normalized to the same PTV dose coverage. CK factors (CF) (defined as modulation scaling factor in this work), number of nodes (NN), number of MLC segments (NS) and beam delivery time (BT) with an estimated image interval of 60 seconds, were used for evaluation of delivery efficiency. Results: Generated plans from the MP and Eclipse TPS demonstrated the similar quality in terms of PTV confomality index, minimum and maximum PTV doses, and doses received by critical structures. Group one: CF ranged from 8.1 to 8.7, NN from 30 to 40, NS from 120 to 155 and BT from 20 to 23 minutes; group two: CF from 4.7 to 8.5, NN from 15 to 19, NS from 82 to 141 and BT from 18 to 24 minutes; and group three: CF from 7.9 to 10, NN from 47 to 49, NS from 110 to 113 and BT from 20 to 22 minutes. Conclusions: Delivery time is longer for the CK M6 than for the Varian linac (7.8 to 13.7 minutes). Further investigation will be necessary to determine if a PTV reduction from the tracking feature will shorten the delivery time without decreasing plan quality.

OSTI ID:
22369666
Journal Information:
Medical Physics, Vol. 41, Issue 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language: