Title: SU-E-T-472: A Multi-Dimensional Measurements Comparison to Analyze a 3D Patient Specific QA Tool

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4888805· OSTI ID:22369615
; ; ; ; ;  [1]
  1. Duke University Medical Center, Durham, NC (United States)

Purpose: To quantitatively evaluate a 3D patient specific QA tool using 2D film and 3D Presage dosimetry. Methods: A brain IMRT case was delivered to Delta4, EBT2 film and Presage plastic dosimeter. The film was inserted in the solid water slabs at 7.5cm depth for measurement. The Presage dosimeter was inserted into a head phantom for 3D dose measurement. Delta4's Anatomy software was used to calculate the corresponding dose to the film in solid water slabs and to Presage in the head phantom. The results from Anatomy were compared to both calculated results from Eclipse and measured dose from film and Presage to evaluate its accuracy. Using RIT software, we compared the “Anatomy” dose to the EBT2 film measurement and the film measurement to ECLIPSE calculation. For 3D analysis, DICOM file of “Anatomy” was extracted and imported to CERR software, which was used to compare the Presage dose to both “Anatomy” calculation and ECLIPSE calculation. Gamma criteria of 3% - 3mm and 5% - 5mm was used for comparison. Results: Gamma passing rates of film vs “Anatomy”, “Anatomy” vs ECLIPSE and film vs ECLIPSE were 82.8%, 70.9% and 87.6% respectively when 3% - 3mm criteria is used. When the criteria is changed to 5% - 5mm, the passing rates became 87.8%, 76.3% and 90.8% respectively. For 3D analysis, Anatomy vs ECLIPSE showed gamma passing rate of 86.4% and 93.3% for 3% - 3mm and 5% - 5mm respectively. The rate is 77.0% for Presage vs ECLIPSE analysis. The Anatomy vs ECLIPSE were absolute dose comparison. However, film and Presage analysis were relative comparison Conclusion: The results show higher passing rate in 3D than 2D in “Anatomy” software. This could be due to the higher degrees of freedom in 3D than in 2D for gamma analysis.

OSTI ID:
22369615
Journal Information:
Medical Physics, Vol. 41, Issue 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English

Similar Records

SU-E-T-341: DVH-Based Comparison of True 3D Measurements to a Delta4 System
Journal Article · Mon Jun 15 00:00:00 EDT 2015 · Medical Physics · OSTI ID:22369615

SU-E-T-159: Evaluation of a Patient Specific QA Tool Based On TG119
Journal Article · Sun Jun 01 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22369615

SU-E-T-535: Preliminary 2D and 3D Gamma Calculation Comparison Using PRESAGE
Journal Article · Sun Jun 01 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22369615