Title: SU-E-T-317: The Development of a DIBH Technique for Left Sided Breast Patients Undergoing Radiation Therapy Utilizing Varians RPM System in a Community Hospital

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4888650· OSTI ID:22355873
; ; ;  [1]
  1. Anne Arundel Medical Center, Annapolis, MD (United States)

Purpose: To develop and implement a Deep Inhalation Breath Hold program (DIBH) for treatment of patients with Left-sided breast cancer in a community hospital. Methods: All patients with left sided breast cancer underwent a screening free breathing CT. Evaluation of the conventional tangent treatment fields and the heart was conducted. If the heart would not be excluded using tangents, the patient then received DIBH breathe coaching. The patients returned for a 4D CT simulation. The patients breathing cycle was monitored using the Varian Real-Time position ManagementTM (RPM) system to assess duration of DIBH, amplitude, phase and recovery time to normal breathing. Then a DIBH CT was obtained at the desired amplitude. Duplicate plans were developed for both free breathing and DIBH on the Eclipse planning system and comparison DVH's were created. The plan that provided the prescribed treatment coverage and the least doses to the OAR (heart, Lt. Lung) was determined. Those patients selected to receive treatment with DIBH were set up for treatment, and breathing was monitored using the RPM system. Practice trials were used to confirm that the amplitude, phase and recovery were consistent with findings from simulation. Results: 10 patients have been treated using the DIBH procedure in our clinic. The DIBH patients had an average increase of 80% lung volume on DIBH, decreased lung volume receiving 50% of the dose, and decreases in the V20 dose. Significant reduction in the maximum and mean dose to the heart, as well as the dose to 1CC of the volume for the DIBH plans. Conclusion: Using the RPM system already available in the clinic, staff training, and patient coaching a simple DIBH program was setup. The use of DIBH has shown promise in reducing doses to the critical organs while maintaining PTV coverage for left sided breast treatments.

OSTI ID:
22355873
Journal Information:
Medical Physics, Vol. 41, Issue 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English