skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Demonstration of improvement in the signal-to-noise ratio of Thomson scattering signal obtained by using a multi-pass optical cavity on the Tokyo Spherical Tokamak-2

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.4891707· OSTI ID:22308929
; ; ; ; ; ; ; ; ; ; ;  [1]; ; ; ;  [2]; ; ;  [3];
  1. Graduate School of Frontier Sciences, The University of Tokyo, Kashiwa 277-8561 (Japan)
  2. Graduate School of Science, The University of Tokyo, Tokyo 113-0033 (Japan)
  3. Research Institute of Applied Mechanics, Kyushu University, Kasuga 816-8580 (Japan)

The multi-pass Thomson scattering (TS) scheme enables obtaining many photons by accumulating multiple TS signals. The signal-to-noise ratio (SNR) depends on the accumulation number. In this study, we performed multi-pass TS measurements for ohmically heated plasmas, and the relationship between SNR and the accumulation number was investigated. As a result, improvement of SNR in this experiment indicated similar tendency to that calculated for the background noise dominant situation.

OSTI ID:
22308929
Journal Information:
Review of Scientific Instruments, Vol. 85, Issue 11; Other Information: (c) 2014 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States
Language:
English