Title: Evaluation of two-stage system for neutron measurement aiming at increase in count rate at Japan Atomic Energy Agency-Fusion Neutronics Source

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.4893941· OSTI ID:22303726
;  [1]; ;  [2];  [3];  [4]
  1. Japan Atomic Energy Agency, Naka, Ibaraki 311-0193 (Japan)
  2. Department of Quantum Science and Energy Engineering, Tohoku University, Sendai, Miyagi 980-8579 (Japan)
  3. Cyclotron and Radioisotope Center, Tohoku University, Sendai, Miyagi 980-8578 (Japan)
  4. Organization for Research Initiatives and Development, Doshisha University, Kyoto 602-8580 (Japan)

In order to increase the count rate capability of a neutron detection system as a whole, we propose a multi-stage neutron detection system. Experiments to test the effectiveness of this concept were carried out on Fusion Neutronics Source. Comparing four configurations of alignment, it was found that the influence of an anterior stage on a posterior stage was negligible for the pulse height distribution. The two-stage system using 25 mm thickness scintillator was about 1.65 times the count rate capability of a single detector system for d-D neutrons and was about 1.8 times the count rate capability for d-T neutrons. The results suggested that the concept of a multi-stage detection system will work in practice.

OSTI ID:
22303726
Journal Information:
Review of Scientific Instruments, Vol. 85, Issue 11; Other Information: (c) 2014 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States