Title: Cost-effective and highly sensitive cholesterol microsensors with fast response based on the enzyme-induced conductivity change of polyaniline

Journal Article · · Applied Physics Letters
DOI:https://doi.org/10.1063/1.4896289· OSTI ID:22303533
; ; ; ; ; ; ; ; ;  [1];  [2];  [3]
  1. Institute of Nanoengineering and Microsystems, National Tsing Hua University, Hsinchu 300, Taiwan (China)
  2. Division of Medical Engineering, National Health Research Institutes, MiaoLi, Taiwan (China)
  3. Department of Biomedical Engineering and Environmental Science, National Tsing Hua University, Hsinchu 300, Taiwan (China)

In this study, a cost-effective and highly sensitive cholesterol microsensor, which is consisted of cholesterol oxidase (ChOx), horseradish peroxidase (HRP), and polyaniline (PANI), was developed based on the enzyme-induced conductivity change of PANI with fast response. Hydrogen peroxide is produced via the reaction between cholesterol and ChOx, which was immobilized in a dialysis membrane. The produced hydrogen peroxide can oxidize HRP, which can be reduced by oxidizing PANI, thus resulting in decreased conductivity of the polyaniline thin film. The reduced HRP can be oxidized again by hydrogen peroxide and the cycle of the oxidation/reduction continues until all hydrogen peroxide are reacted, leading to the high sensitivity of the sensor due to the signal contributed from all hydrogen peroxide molecules. Cholesterol was detected near the physiological concentrations ranging from 100 mg/dl to 400 mg/dl with the cholesterol microsensors. The results show linear relation between cholesterol concentration and the conductivity change of the PANI. The microsensor showed no response to cholesterol when the PANI was standalone without cholesterol oxidase immobilized, indicating that the enzymatic reaction is required for cholesterol detection. The simple process of the sensor fabrication allows the sensor to be cost-effective and disposable usage. This electronic cholesterol microsensor is promising for point-of-care health monitoring in cholesterol level with low cost and fast response.

OSTI ID:
22303533
Journal Information:
Applied Physics Letters, Vol. 105, Issue 11; Other Information: (c) 2014 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0003-6951
Country of Publication:
United States
Language:
English