Use of proton beams with breast prostheses and tissue expanders
- ProCure Proton Therapy Center, Somerset, NJ (United States)
- ProCure Proton Therapy Center, Warrenville, IL (United States)
Since the early 2000s, a small but rapidly increasing number of patients with breast cancer have been treated with proton beams. Some of these patients have had breast prostheses or tissue expanders in place during their courses of treatment. Procedures must be implemented to plan the treatments of these patients. The density, kilovoltage x-ray computed tomography numbers (kVXCTNs), and proton relative linear stopping powers (pRLSPs) were calculated and measured for several test sample devices. The calculated and measured kVXCTNs of saline were 1% and 2.4% higher than the values for distilled water while the calculated RLSP for saline was within 0.2% of the value for distilled water. The measured kVXCTN and pRLSP of the silicone filling material for the test samples were approximately 1120 and 0.935, respectively. The conversion of kVXCTNs to pRLSPs by the treatment planning system standard tissue conversion function is adequate for saline-filled devices but for silicone-filled devices manual reassignment of the pRLSPs is required.
- OSTI ID:
- 22267977
- Journal Information:
- Medical Dosimetry, Vol. 39, Issue 1; Other Information: Copyright (c) 2014 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.; Country of input: International Atomic Energy Agency (IAEA); ISSN 0958-3947
- Country of Publication:
- United States
- Language:
- English
Similar Records
Changes in photon dose distributions due to breast prostheses
SU-F-T-409: Modelling of the Magnetic Port in Temporary Breast Tissue Expanders for a Treatment Planning System