Title: Cumulative dose on fractional delivery of tomotherapy to periodically moving organ: A phantom QA suggestion

Journal Article · · Medical Dosimetry
 [1];  [1];  [1]; ; ; ; ;  [1];  [2]
  1. Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul (Korea, Republic of)
  2. Department of Nuclear Engineering, Hanyang University, Seoul (Korea, Republic of)

This study was conducted to evaluate the cumulative dosimetric error that occurs in both target and surrounding normal tissues when treating a moving target in multifractional treatment with tomotherapy. An experiment was devised to measure cumulative error in multifractional treatments delivered to a horseshoe-shaped clinical target volume (CTV) surrounding a cylinder shape of organ at risk (OAR). Treatments differed in jaw size (1.05 vs 2.5 cm), pitch (0.287 vs 0.660), and modulation factor (1.5 vs 2.5), and tumor motion characteristics differing in amplitude (1 to 3 cm), period (3 to 5 second), and regularity (sinusoidal vs irregular) were tested. Treatment plans were delivered to a moving phantom up to 5-times exposure. Dose distribution on central coronal plane from 1 to 5 times exposure was measured with GAFCHROMIC EBT film. Dose differences occurring across 1 to 5 times exposure of treatment and between treatment plans were evaluated by analyzing measurements of gamma index, gamma index histogram, histogram changes, and dose at the center of the OAR. The experiment showed dose distortion due to organ motion increased between multiexposure 1 to 3 times but plateaued and remained constant after 3-times exposure. In addition, although larger motion amplitude and a longer period of motion both increased dosimetric error, the dose at the OAR was more significantly affected by motion amplitude rather than motion period. Irregularity of motion did not contribute significantly to dosimetric error when compared with other motion parameters. Restriction of organ motion to have small amplitude and short motion period together with larger jaw size and small modulation factor (with small pitch) is effective in reducing dosimetric error. Pretreatment measurements for 3-times exposure of treatment to a moving phantom with patient-specific tumor motion would provide a good estimation of the delivered dose distribution.

OSTI ID:
22262835
Journal Information:
Medical Dosimetry, Vol. 38, Issue 4; Other Information: Copyright (c) 2013 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.; Country of input: International Atomic Energy Agency (IAEA); ISSN 0958-3947
Country of Publication:
United States
Language:
English