Title: Free software for performing physical analysis of systems for digital radiography and mammography

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4870955· OSTI ID:22250710
;  [1];  [2];  [3]
  1. Alma Mater Studiorum, Department of Physics and Astronomy, University of Bologna, Bologna 40127 (Italy)
  2. Fisica Medica, Ospedale di Sassuolo S.p.A., Sassuolo 41049 (Italy)
  3. Medical Physics Unit, Azienda Ospedaliera ASMN, Istituto di Ricovero e Cura a Carattere Scientifico, Reggio Emilia 42123 (Italy)

Purpose: In this paper, the authors present a free software for assisting users in achieving the physical characterization of x-ray digital systems and image quality checks. Methods: The program was developed as a plugin of a well-known public-domain suite ImageJ. The software can assist users in calculating various physical parameters such as the response curve (also termed signal transfer property), modulation transfer function (MTF), noise power spectra (NPS), and detective quantum efficiency (DQE). It also includes the computation of some image quality checks: defective pixel analysis, uniformity, dark analysis, and lag. Results: The software was made available in 2009 and has been used during the last couple of years by many users who gave us valuable feedback for improving its usability. It was tested for achieving the physical characterization of several clinical systems for digital radiography and mammography. Various published papers made use of the outcomes of the plugin. Conclusions: This software is potentially beneficial to a variety of users: physicists working in hospitals, staff working in radiological departments, such as medical physicists, physicians, engineers. The plugin, together with a brief user manual, are freely available and can be found online ( http://www.medphys.it/downloads.htm ). With our plugin users can estimate all three most important parameters used for physical characterization (MTF, NPS, and also DQE). The plugin can run on any operating system equipped with ImageJ suite. The authors validated the software by comparing MTF and NPS curves on a common set of images with those obtained with other dedicated programs, achieving a very good agreement.

OSTI ID:
22250710
Journal Information:
Medical Physics, Vol. 41, Issue 5; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English