Time-resolved measurement of plasma parameters by means of triple probe
- National Tokamak Fusion Program, 3329 Islamabad (Pakistan)
Triple Langmuir probe (TLP) diagnostic system with its necessary driving circuit is developed and successfully applies for time-resolved measurement of plasma parameters in the negative glow region of pulsating-dc discharge. This technique allows the instantaneous measurement of electron temperature [T{sub −}], electron number density [n{sub −}] as well as plasma fluctuations without any voltage or frequency sweep. In TLP configuration two probes are differentially biased and serve as a floating symmetric double probe whereas the third probe is simply floating into plasma to measure floating potential as a function of time and thus incorporates the effect of plasma fluctuations. As an example of the application to time-dependent plasmas, basic plasma parameters such as floating potential, electron temperature, and electron number density in low pressure air discharge are determined as a function of time for different fill pressure. The results demonstrate temporal evolution of plasma parameters and thus plasma generation progression for different fill pressures.
- OSTI ID:
- 22220216
- Journal Information:
- Review of Scientific Instruments, Vol. 84, Issue 12; Other Information: (c) 2013 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
- Country of Publication:
- United States
- Language:
- English
Similar Records
Practical solutions for reliable triple probe measurements in magnetized plasmas
Time-resolved triple probe investigations of a pulsed magnetron discharge