skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Improvement of Stent Retriever Design and Efficacy of Mechanical Thrombectomy in a Flow Model

Journal Article · · Cardiovascular and Interventional Radiology
 [1];  [2]; ;  [1]
  1. Institute of Neuroradiology, University of Frankfurt am Main (Germany)
  2. Acandis GmbH and Co KG (Germany)

In vitro experiments were performed to evaluate the efficacy of mechanical intracranial thrombectomy comparing the newly developed Aperio stent retriever and standard devices for stroke treatment. The Aperio (A), with an increased working length of 4 cm and a special cell design for capturing and withholding clots, was compared to three benchmark devices: the Solitaire retrievable stent (B), the Merci X6 (C), and the Merci L5 retriever (D). In a vascular glass model with pulsatile flow, reminiscent of the M1 segment of the middle cerebral artery, we repeatedly induced occlusion by generating thrombi via a modified Chandler loop system. The numbers of recanalization attempts, peripheral embolizations, and recanalizations at the site of occlusion were recorded during 10 retrieval experiments with each device. Eleven devices were able to remove the blood clots from the occluded branch. In 34 of 40 experiments, restoration of flow was obtained in 1-3 attempts. The main differences between the study devices were observed in terms of clot withholding and fragmentation during retrieval. Although there was only one fragmentation recorded for device A, disengagement of the whole clot or peripheral embolization of fragments occurred more frequently (5-7 times) with devices B, C, and D. In a vascular model, the design of device A was best at capturing and withholding thrombi during retrieval. Further study will be necessary to see whether this holds true in clinical applications.

OSTI ID:
22163098
Journal Information:
Cardiovascular and Interventional Radiology, Vol. 36, Issue 1; Other Information: Copyright (c) 2013 Springer Science+Business Media New York and the Cardiovascular and Interventional Radiological Society of Europe (CIRSE); Article Copyright (c) 2012 Springer Science+Business Media, LLC and the Cardiovascular and Interventional Radiological Society of Europe (CIRSE); Country of input: International Atomic Energy Agency (IAEA); ISSN 0174-1551
Country of Publication:
United States
Language:
English

Similar Records

A Novel Mechanical Thrombectomy Device for Retrieval of Intravascular Thrombus
Journal Article · Fri Apr 15 00:00:00 EDT 2011 · Cardiovascular and Interventional Radiology · OSTI ID:22163098

Influence of Thrombus Age on the Mechanical Thrombectomy Efficacy of the Amplatz Thrombectomy Device In Vitro
Journal Article · Sun Jun 15 00:00:00 EDT 2003 · Cardiovascular and Interventional Radiology · OSTI ID:22163098

Recanalization of Acute and Subacute Femoropopliteal Artery Occlusions with the Rotarex Catheter: One Year Follow-up, Single Center Experience
Journal Article · Wed Jun 15 00:00:00 EDT 2005 · Cardiovascular and Interventional Radiology · OSTI ID:22163098