skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Note: Readout of a micromechanical magnetometer for the ITER fusion reactor

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.4807718· OSTI ID:22118570
;  [1]
  1. VTT Technical Research Centre of Finland, Tietotie 3, 02150 Espoo (Finland)

We present readout instrumentation for a MEMS magnetometer, placed 30 m away from the MEMS element. This is particularly useful when sensing is performed in high-radiation environment, where the semiconductors in the readout cannot survive. High bandwidth transimpedance amplifiers are used to cancel the cable capacitances of several nanofarads. A frequency doubling readout scheme is used for crosstalk elimination. Signal-to-noise ratio in the range of 60 dB was achieved and with sub-percent nonlinearity. The presented instrument is intended for the steady-state magnetic field measurements in the ITER fusion reactor.

OSTI ID:
22118570
Journal Information:
Review of Scientific Instruments, Vol. 84, Issue 5; Other Information: (c) 2013 Euratom; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States
Language:
English