Development of a time-variable nuclear pulser for half life measurements
Abstract
In this work a time-variable pulser system with an exponentially-decaying pulse frequency is presented, which was developed using the low-cost, open-source Arduino microcontroler plataform. In this system, the microcontroller produces a TTL signal in the selected rate and a pulse shaper board adjusts it to be entered in an amplifier as a conventional pulser signal; both the decay constant and the initial pulse rate can be adjusted using a user-friendly control software, and the pulse amplitude can be adjusted using a potentiometer in the pulse shaper board. The pulser was tested using several combinations of initial pulse rate and decay constant, and the results show that the system is stable and reliable, and is suitable to be used in half-life measurements.
- Authors:
-
- Instituto de Pesquisas Energeticas e Nucleares - IPEN-CNEN/SP. P.O. Box 11049, Sao Paulo, 05422-970 (Brazil)
- Publication Date:
- OSTI Identifier:
- 22111940
- Resource Type:
- Journal Article
- Journal Name:
- AIP Conference Proceedings
- Additional Journal Information:
- Journal Volume: 1529; Journal Issue: 1; Conference: 35. Brazilian workshop on nuclear physics, Sao Paulo (Brazil), 2-6 Sep 2012; Other Information: (c) 2013 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA); Journal ID: ISSN 0094-243X
- Country of Publication:
- United States
- Language:
- English
- Subject:
- 46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; AMPLIFIERS; COMPUTER CODES; HALF-LIFE; NUCLEAR DECAY; POTENTIOMETERS; PULSE SHAPERS; PULSES; RADIATION DETECTION; RADIATION DETECTORS; TIME DEPENDENCE; TRANSISTORS
Citation Formats
Zahn, Guilherme S., Domienikan, Claudio, Carvalhaes, Roberto P. M., and Genezini, Frederico A. Development of a time-variable nuclear pulser for half life measurements. United States: N. p., 2013.
Web. doi:10.1063/1.4804106.
Zahn, Guilherme S., Domienikan, Claudio, Carvalhaes, Roberto P. M., & Genezini, Frederico A. Development of a time-variable nuclear pulser for half life measurements. United States. https://doi.org/10.1063/1.4804106
Zahn, Guilherme S., Domienikan, Claudio, Carvalhaes, Roberto P. M., and Genezini, Frederico A. 2013.
"Development of a time-variable nuclear pulser for half life measurements". United States. https://doi.org/10.1063/1.4804106.
@article{osti_22111940,
title = {Development of a time-variable nuclear pulser for half life measurements},
author = {Zahn, Guilherme S. and Domienikan, Claudio and Carvalhaes, Roberto P. M. and Genezini, Frederico A.},
abstractNote = {In this work a time-variable pulser system with an exponentially-decaying pulse frequency is presented, which was developed using the low-cost, open-source Arduino microcontroler plataform. In this system, the microcontroller produces a TTL signal in the selected rate and a pulse shaper board adjusts it to be entered in an amplifier as a conventional pulser signal; both the decay constant and the initial pulse rate can be adjusted using a user-friendly control software, and the pulse amplitude can be adjusted using a potentiometer in the pulse shaper board. The pulser was tested using several combinations of initial pulse rate and decay constant, and the results show that the system is stable and reliable, and is suitable to be used in half-life measurements.},
doi = {10.1063/1.4804106},
url = {https://www.osti.gov/biblio/22111940},
journal = {AIP Conference Proceedings},
issn = {0094-243X},
number = 1,
volume = 1529,
place = {United States},
year = {Mon May 06 00:00:00 EDT 2013},
month = {Mon May 06 00:00:00 EDT 2013}
}