Title: Performance characterization of a MVCT scanner using multislice thick, segmented cadmium tungstate-photodiode detectors

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.3273032· OSTI ID:22102180
; ; ;  [1]
  1. Department of Physics, University of Alberta, Edmonton, Alberta T6E 2E1 (Canada)

Purpose: Megavoltage computed tomography (MVCT) and megavoltage cone beam computed tomography (MVCBCT) can be used for visualizing anatomical structures prior to radiation therapy treatments to assist in patient setup and target localization. These systems are less susceptible to metal artifacts and provide better CT number linearity than conventional CT scanners. However, their contrast is limited by the properties of the megavoltage photons and the low detective quantum efficiency (DQE) of flat panel detector systems currently available. By using higher DQE, thick, segmented cadmium tungstate detectors, the authors can improve the low contrast detectability of a MVCT system. This in turn would permit greater soft tissue visualization for a given radiation dose, allowing MVCT to be used in more clinical situations. Methods: This article describes the evaluation of our prototype system that uses thick, segmented detectors. In order to create images using a dose that would be acceptable for day to day patient imaging, the authors evaluated their system using the low intensity bremsstrahlung component of a 6 MeV electron beam. The system was evaluated for its uniformity, high contrast resolution, low contrast detectability, signal to noise ratio, contrast to noise ratio, and CT number linearity. Results: The prototype system was found to have a high contrast spatial resolution of about 5 line pairs per cm, and to be able to visualize a 15 mm 1.5% contrast target with 2 cGy of radiation dose delivered. SNR{sup 2} vs radiation dose and mean pixel value vs electron density curves were linear. Conclusions: This prototype system shows a large improvement in low contrast detectability over current MVCBCT systems.

OSTI ID:
22102180
Journal Information:
Medical Physics, Vol. 37, Issue 1; Other Information: (c) 2010 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English