Title: Effect of jaw size in megavoltage CT on image quality and dose

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4736951· OSTI ID:22098952
; ; ; ; ; ; ; ; ; ; ;  [1]
  1. Department of Radiation Oncology, College of Medicine, Soonchunhyang University Bucheon Hospital, Bucheon 1174, Korea and Department of Biomedical Engineering and Research Institute of Biomedical Engineering, College of Medicine, Catholic University of Korea Seoul 137-701 (Korea, Republic of)

Purpose: Recently, the jaw size for the TomoTherapy Hi-Art II{sup Registered-Sign} (TomoTherapy Inc., Madison, WI) was reduced from 4 mm (J4) to 1 mm (J1) to improve the longitudinal (IEC-Y) resolution in megavoltage computed tomography (MVCT) images. This study evaluated the effect of jaw size on the image quality and dose, as well as the dose delivered to the lens of the eye, which is a highly radiosensitive tissue. Methods: MVCT image quality (image noise, uniformity, contrast linearity, high-contrast resolution, and full width at half-maximum) and multiple scan average dose (MSAD) were measured at different jaw sizes. A head phantom and photoluminescence glass dosimeters (PLDs) were used to measure the exposed lens dose (cGy). Different MVCT scan modes (pitch = 1, 2, and 3) and scan lengths (108 mm, 156 mm, and 204 mm) were applied in the MSAD and PLDs measurements. Results: The change in jaw size from J4 to J1 produced no change or only a slight improvement in image noise, uniformity, contrast linearity, and high-contrast resolution. However, the full-width at half-maximum reduced from approximately 7.2 at J4 to 4.5 mm at J1, which represents an enhancement in the longitudinal resolution. The MSAD at the center point changed from approximately 0.69-2.32 cGy (peripheral: 0.83-2.49 cGy) at J4 to 0.85-2.81 cGy (peripheral: 1.05-2.86 cGy) at J1. The measured lens dose increased from 0.92-3.36 cGy at J4 to 1.06-3.91 cGy at J1. Conclusions: The change in jaw size improved longitudinal resolution. The MVCT imaging dose of approximately 3.86 cGy, 1.92 cGy, and 1.22 cGy was delivered at a pitch of 1, 2, and 3, respectively, per fraction in the head and neck treatment plans. Therefore, allowance for an approximately 15% increase in lens dose over that with J4 should be provided with J1.

OSTI ID:
22098952
Journal Information:
Medical Physics, Vol. 39, Issue 8; Other Information: (c) 2012 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English

Similar Records

Quantitative characterization of tomotherapy MVCT dosimetry
Journal Article · Tue Oct 01 00:00:00 EDT 2013 · Medical Dosimetry · OSTI ID:22098952

Patient Dose From Megavoltage Computed Tomography Imaging
Journal Article · Tue Apr 01 00:00:00 EDT 2008 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:22098952

High-dose MVCT image guidance for stereotactic body radiation therapy
Journal Article · Wed Aug 15 00:00:00 EDT 2012 · Medical Physics · OSTI ID:22098952