skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Imaging of moving fiducial markers during radiotherapy using a fast, efficient active pixel sensor based EPID

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.3651632· OSTI ID:22098673
; ; ; ; ;  [1]
  1. Joint Department of Physics, Institute of Cancer Research and The Royal Marsden NHS Foundation Trust, Downs Road, Sutton, Surrey, SM2 5PT (United Kingdom)

Purpose: The purpose of this work was to investigate the use of an experimental complementary metal-oxide-semiconductor (CMOS) active pixel sensor (APS) for tracking of moving fiducial markers during radiotherapy. Methods: The APS has an active area of 5.4 x 5.4 cm and maximum full frame read-out rate of 20 frame s{sup -1}, with the option to read out a region-of-interest (ROI) at an increased rate. It was coupled to a 4 mm thick ZnWO4 scintillator which provided a quantum efficiency (QE) of 8% for a 6 MV x-ray treatment beam. The APS was compared with a standard iViewGT flat panel amorphous Silicon (a-Si) electronic portal imaging device (EPID), with a QE of 0.34% and a frame-rate of 2.5 frame s{sup -1}. To investigate the ability of the two systems to image markers, four gold cylinders of length 8 mm and diameter 0.8, 1.2, 1.6, and 2 mm were placed on a motion-platform. Images of the stationary markers were acquired using the APS at a frame-rate of 20 frame s{sup -1}, and a dose-rate of 143 MU min{sup -1} to avoid saturation. EPID images were acquired at the maximum frame-rate of 2.5 frame s{sup -1}, and a reduced dose-rate of 19 MU min{sup -1} to provide a similar dose per frame to the APS. Signal-to-noise ratio (SNR) of the background signal and contrast-to-noise ratio (CNR) of the marker signal relative to the background were evaluated for both imagers at doses of 0.125 to 2 MU. Results: Image quality and marker visibility was found to be greater in the APS with SNR {approx}5 times greater than in the EPID and CNR up to an order of magnitude greater for all four markers. To investigate the ability to image and track moving markers the motion-platform was moved to simulate a breathing cycle with period 6 s, amplitude 20 mm and maximum speed 13.2 mm s{sup -1}. At the minimum integration time of 50 ms a tracking algorithm applied to the APS data found all four markers with a success rate of {>=}92% and positional error {<=}90 {mu}m. At an integration time of 400 ms the smallest marker became difficult to detect when moving. The detection of moving markers using the a-Si EPID was difficult even at the maximum dose-rate of 592 MU min{sup -1} due to the lower QE and longer integration time of 400 ms. Conclusions: This work demonstrates that a fast read-out, high QE APS may be useful in the tracking of moving fiducial markers during radiotherapy. Further study is required to investigate the tracking of markers moving in 3D in a treatment beam attenuated by moving patient anatomy. This will require a larger sensor with ROI read-out to maintain speed and a manageable data-rate.

OSTI ID:
22098673
Journal Information:
Medical Physics, Vol. 38, Issue 11; Other Information: (c) 2011 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English