Title: Wide field array calibration dependence on the stability of measured dose distributions

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.3442028· OSTI ID:22096710
; ; ; ;  [1]
  1. Department of Nuclear and Radiological Engineering, University of Florida, 202 Nuclear Science Building, Gainesville, Florida 32611-8300, Department of Radiation Oncology, Health Science Center, University of Florida, P.O. Box 100385, Gainesville, Florida 32610-0385, and Sun Nuclear Corporation, 425-A Pineda Court, Melbourne, Florida 32940 (United States)

Purpose: The aim of this work was to simulate the effect of dose distribution changes on detector array calibrations and to explore compensatory methods that are used during calibration measurements. Methods: The array calibration technique that was investigated is known as wide field (WF) calibration. Using this method, a linear array [y-axis (65 detectors) of the IC PROFILER (Sun Nuclear Corporation, Melbourne, FL)] is calibrated with three measurements ({alpha}, {theta}, and {lambda}); each measurement uses the same radiation field, which is larger than the array. For measurement configuration {theta}, the array is rotated by 180 deg. from its position in {alpha}; for {lambda}, the array is shifted by one detector from its position in {theta}. The relative detector sensitivities are then determined through ratios of detector readings at the same field locations (using {theta} and {lambda}). This method results in error propagation that is proportional to the number of detectors in the array. During the procedure, the calibration protocol operates under three postulates, which state that (a) the beam shape does not change between measurements; (b) the relative sensitivities of the detectors do not change; and (c) the scatter to the array does not change as the array is moved. The WF calibration's sensitivity to a postulate (a) violation was quantified by applying a sine shaped perturbation (of up to 0.1%) to {alpha}, {theta}, or {lambda}, and then determining the change relative to a baseline calibration. Postulate (a) violations were minimized by using a continuous beam and mechanized array movement during {theta} and {lambda}. A continuously on beam demonstrated more stable beam symmetry as compared to cycling the beam on and off between measurements. Additional side-scatter was also used to satisfy postulate (c). Results: Simulated symmetry perturbations of 0.1% to {theta} or {lambda} resulted in calibration errors of up to 2%; {alpha} was relatively immune to perturbation (<0.1% error). Wide field calibration error on a linear accelerator with similar symmetry variations was {+-}1.6%. Using a continuous beam during {theta} and {lambda} with additional side-scatter reduced the calibration error from {+-}1.6% to {+-}0.48%. Conclusions: This work increased the reproducibility of WF calibrations by limiting the effect of measurement perturbations primarily due to linear accelerator symmetry variations. The same technique would work for any array using WF calibration.

OSTI ID:
22096710
Journal Information:
Medical Physics, Vol. 37, Issue 7; Other Information: (c) 2010 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English