Title: Real-time tumor tracking: Automatic compensation of target motion using the Siemens 160 MLC

Abstract

Purpose: Advanced high quality radiation therapy techniques such as IMRT require an accurate delivery of precisely modulated radiation fields to the target volume. Interfractional and intrafractional motion of the patient's anatomy, however, may considerably deteriorate the accuracy of the delivered dose to the planned dose distributions. In order to compensate for these potential errors, a dynamic real-time capable MLC control system was designed. Methods: The newly developed adaptive MLC control system contains specialized algorithms which are capable of continuous optimization and correction of the aperture of the MLC according to the motion of the target volume during the dose delivery. The algorithms calculate the new leaf positions based on target information provided online to the system. The algorithms were implemented in a dynamic target tracking control system designed for a Siemens 160 MLC. To assess the quality of the new target tracking system in terms of dosimetric accuracy, experiments with various types of motion patterns using different phantom setups were performed. The phantoms were equipped with radiochromic films placed between solid water slabs. Dosimetric results of exemplary deliveries to moving targets with and without dynamic MLC tracking applied were compared in terms of the gamma criterion to the reference dosemore » delivered to a static phantom. Results: Our measurements indicated that dose errors for clinically relevant two-dimensional target motion can be compensated by the new control system during the dose delivery of open fields. For a clinical IMRT dose distribution, the gamma success rate was increased from 19% to 77% using the new tracking system. Similar improvements were achieved for the delivery of a complete IMRT treatment fraction to a moving lung phantom. However, dosimetric accuracy was limited by the system's latency of 400 ms and the finite leaf width of 5 mm in the isocenter plane. Conclusions: Different experimental setups representing different target tracking scenarios proved that the tracking concept, the new algorithms and the dynamic control system make it possible to effectively compensate for dose errors due to target motion in real-time. These early results indicate that the method is suited to increasing the accuracy and the quality of the treatment delivery for the irradiation of moving tumors.« less

Authors:
; ; ;  [1]
  1. Department of Medical Physics in Radiation Oncology, German Cancer Research Center, Im Neuenheimer Feld 280, D-69120 Heidelberg (Germany)
Publication Date:
OSTI Identifier:
22096632
Resource Type:
Journal Article
Journal Name:
Medical Physics
Additional Journal Information:
Journal Volume: 37; Journal Issue: 2; Other Information: (c) 2010 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); Journal ID: ISSN 0094-2405
Country of Publication:
United States
Language:
English
Subject:
62 RADIOLOGY AND NUCLEAR MEDICINE; 61 RADIATION PROTECTION AND DOSIMETRY; ACCURACY; ALGORITHMS; CONTROL SYSTEMS; DOSIMETRY; FILMS; IMAGE PROCESSING; IRRADIATION; LUNGS; NEOPLASMS; OPTIMIZATION; PATIENTS; PHANTOMS; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; RADIOTHERAPY

Citation Formats

Tacke, Martin B., Nill, Simeon, Krauss, Andreas, and Oelfke, Uwe. Real-time tumor tracking: Automatic compensation of target motion using the Siemens 160 MLC. United States: N. p., 2010. Web. doi:10.1118/1.3284543.
Tacke, Martin B., Nill, Simeon, Krauss, Andreas, & Oelfke, Uwe. Real-time tumor tracking: Automatic compensation of target motion using the Siemens 160 MLC. United States. https://doi.org/10.1118/1.3284543
Tacke, Martin B., Nill, Simeon, Krauss, Andreas, and Oelfke, Uwe. 2010. "Real-time tumor tracking: Automatic compensation of target motion using the Siemens 160 MLC". United States. https://doi.org/10.1118/1.3284543.
@article{osti_22096632,
title = {Real-time tumor tracking: Automatic compensation of target motion using the Siemens 160 MLC},
author = {Tacke, Martin B. and Nill, Simeon and Krauss, Andreas and Oelfke, Uwe},
abstractNote = {Purpose: Advanced high quality radiation therapy techniques such as IMRT require an accurate delivery of precisely modulated radiation fields to the target volume. Interfractional and intrafractional motion of the patient's anatomy, however, may considerably deteriorate the accuracy of the delivered dose to the planned dose distributions. In order to compensate for these potential errors, a dynamic real-time capable MLC control system was designed. Methods: The newly developed adaptive MLC control system contains specialized algorithms which are capable of continuous optimization and correction of the aperture of the MLC according to the motion of the target volume during the dose delivery. The algorithms calculate the new leaf positions based on target information provided online to the system. The algorithms were implemented in a dynamic target tracking control system designed for a Siemens 160 MLC. To assess the quality of the new target tracking system in terms of dosimetric accuracy, experiments with various types of motion patterns using different phantom setups were performed. The phantoms were equipped with radiochromic films placed between solid water slabs. Dosimetric results of exemplary deliveries to moving targets with and without dynamic MLC tracking applied were compared in terms of the gamma criterion to the reference dose delivered to a static phantom. Results: Our measurements indicated that dose errors for clinically relevant two-dimensional target motion can be compensated by the new control system during the dose delivery of open fields. For a clinical IMRT dose distribution, the gamma success rate was increased from 19% to 77% using the new tracking system. Similar improvements were achieved for the delivery of a complete IMRT treatment fraction to a moving lung phantom. However, dosimetric accuracy was limited by the system's latency of 400 ms and the finite leaf width of 5 mm in the isocenter plane. Conclusions: Different experimental setups representing different target tracking scenarios proved that the tracking concept, the new algorithms and the dynamic control system make it possible to effectively compensate for dose errors due to target motion in real-time. These early results indicate that the method is suited to increasing the accuracy and the quality of the treatment delivery for the irradiation of moving tumors.},
doi = {10.1118/1.3284543},
url = {https://www.osti.gov/biblio/22096632}, journal = {Medical Physics},
issn = {0094-2405},
number = 2,
volume = 37,
place = {United States},
year = {Mon Feb 15 00:00:00 EST 2010},
month = {Mon Feb 15 00:00:00 EST 2010}
}