Identification of errors in the electron density measurements of a tangential interferometer/polarimeter system during a tokamak discharge
Journal Article
·
· Review of Scientific Instruments
- Japan Atomic Energy Agency, 801-1, Mukoyama, Naka 311-0193 (Japan)
A new method for the comparative verification of electron density measurements obtained with a tangential interferometer and a polarimeter during a discharge is proposed. The possible errors associated with the interferometer and polarimeter are classified by the time required for their identification. Based on the characteristics of the errors, the fringe shift error of the interferometer and the low-frequency noise of the polarimeter were identified and corrected for the JT-60U tangential interferometer/polarimeter system.
- OSTI ID:
- 22093942
- Journal Information:
- Review of Scientific Instruments, Vol. 83, Issue 10; Other Information: (c) 2012 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
- Country of Publication:
- United States
- Language:
- English
Similar Records
Conceptual design of the tangentially viewing combined interferometer-polarimeter for ITER density measurements
Toroidal interferometer/polarimeter density measurement system on ITER
Dual CO{sub 2} laser interferometer with a wavelength combination of 10.6 and 9.27 {mu}m for electron density measurement on large tokamaks
Journal Article
·
Mon Apr 15 00:00:00 EDT 2013
· Review of Scientific Instruments
·
OSTI ID:22093942
+8 more
Toroidal interferometer/polarimeter density measurement system on ITER
Journal Article
·
Fri Oct 01 00:00:00 EDT 2004
· Review of Scientific Instruments
·
OSTI ID:22093942
+3 more
Dual CO{sub 2} laser interferometer with a wavelength combination of 10.6 and 9.27 {mu}m for electron density measurement on large tokamaks
Journal Article
·
Mon Apr 01 00:00:00 EST 1996
· Review of Scientific Instruments
·
OSTI ID:22093942
+1 more