A new compact, high sensitivity neutron imaging system
- CEA, DAM, DIF,F-91297 Arpajon (France)
- Laboratory for Laser Energetics, University of Rochester, 250 East River Road, Rochester, New York 14623-1299 (United States)
- Lawrence Livermore National Laboratory, Livermore, California 94550 (United States)
We have developed a new small neutron imaging system (SNIS) diagnostic for the OMEGA laser facility. The SNIS uses a penumbral coded aperture and has been designed to record images from low yield (10{sup 9}-10{sup 10} neutrons) implosions such as those using deuterium as the fuel. This camera was tested at OMEGA in 2009 on a rugby hohlraum energetics experiment where it recorded an image at a yield of 1.4 Multiplication-Sign 10{sup 10}. The resolution of this image was 54 {mu}m and the camera was located only 4 meters from target chamber centre. We recently improved the instrument by adding a cooled CCD camera. The sensitivity of the new camera has been fully characterized using a linear accelerator and a {sup 60}Co {gamma}-ray source. The calibration showed that the signal-to-noise ratio could be improved by using raw binning detection.
- OSTI ID:
- 22093911
- Journal Information:
- Review of Scientific Instruments, Vol. 83, Issue 10; Other Information: (c) 2012 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
- Country of Publication:
- United States
- Language:
- English
Similar Records
Experimental Demonstration of X-Ray Drive Enhancement with Rugby-Shaped Hohlraums
Rugby-like hohlraum experimental designs for demonstrating x-ray drive enhancement