Title: Investigating the performance of an ion luminescence probe as a multichannel fast-ion energy spectrometer using pulse height analysis

Abstract

We investigate the capability of a fast-ion luminescent probe to operate as a pulse height ion energy analyzer. An existing high sensitivity system has been reconfigured as a single channel ion detector with an amplifier to give a bandwidth comparable to the phosphor response time. A digital pulse processing method has been developed to determine pulse heights from the detector signal so as to obtain time-resolved information on the ion energy distribution of the plasma ions lost to the wall of the TJ-II stellarator. Finally, the potential of this approach for magnetic confined fusion plasmas is evaluated by studying representative TJ-II discharges.

Authors:
;  [1];  [2]; ;  [3]
  1. Laboratorio Nacional de Fusion, Asociacion Euratom-CIEMAT, Madrid (Spain)
  2. Centro de Microanalisis de Materiales, Universidad Autonoma de Madrid, Madrid (Spain)
  3. Laboratorio de Metrologia de Radiaciones lonizantes, CIEMAT, Madrid (Spain)
Publication Date:
OSTI Identifier:
22093803
Resource Type:
Journal Article
Journal Name:
Review of Scientific Instruments
Additional Journal Information:
Journal Volume: 83; Journal Issue: 10; Other Information: (c) 2012 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); Journal ID: ISSN 0034-6748
Country of Publication:
United States
Language:
English
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; 75 CONDENSED MATTER PHYSICS, SUPERCONDUCTIVITY AND SUPERFLUIDITY; AMPLIFIERS; ENERGY SPECTRA; IONS; LUMINESCENCE; PERFORMANCE; PHOSPHORS; PLASMA CONFINEMENT; PLASMA DIAGNOSTICS; PULSES; SENSITIVITY; SPECTROMETERS; STELLARATORS; TIME RESOLUTION

Citation Formats

Zurro, B., Baciero, A., Jimenez-Rey, D., Rodriguez-Barquero, L., and Crespo, M. T. Investigating the performance of an ion luminescence probe as a multichannel fast-ion energy spectrometer using pulse height analysis. United States: N. p., 2012. Web. doi:10.1063/1.4733572.
Zurro, B., Baciero, A., Jimenez-Rey, D., Rodriguez-Barquero, L., & Crespo, M. T. Investigating the performance of an ion luminescence probe as a multichannel fast-ion energy spectrometer using pulse height analysis. United States. https://doi.org/10.1063/1.4733572
Zurro, B., Baciero, A., Jimenez-Rey, D., Rodriguez-Barquero, L., and Crespo, M. T. 2012. "Investigating the performance of an ion luminescence probe as a multichannel fast-ion energy spectrometer using pulse height analysis". United States. https://doi.org/10.1063/1.4733572.
@article{osti_22093803,
title = {Investigating the performance of an ion luminescence probe as a multichannel fast-ion energy spectrometer using pulse height analysis},
author = {Zurro, B. and Baciero, A. and Jimenez-Rey, D. and Rodriguez-Barquero, L. and Crespo, M. T.},
abstractNote = {We investigate the capability of a fast-ion luminescent probe to operate as a pulse height ion energy analyzer. An existing high sensitivity system has been reconfigured as a single channel ion detector with an amplifier to give a bandwidth comparable to the phosphor response time. A digital pulse processing method has been developed to determine pulse heights from the detector signal so as to obtain time-resolved information on the ion energy distribution of the plasma ions lost to the wall of the TJ-II stellarator. Finally, the potential of this approach for magnetic confined fusion plasmas is evaluated by studying representative TJ-II discharges.},
doi = {10.1063/1.4733572},
url = {https://www.osti.gov/biblio/22093803}, journal = {Review of Scientific Instruments},
issn = {0034-6748},
number = 10,
volume = 83,
place = {United States},
year = {Mon Oct 15 00:00:00 EDT 2012},
month = {Mon Oct 15 00:00:00 EDT 2012}
}