CR mammography: Design and implementation of a quality control program
Journal Article
·
· AIP Conference Proceedings
- Instituto de Fisica, Universidad Nacional Autonoma de Mexico, DF 04510 (Mexico)
Despite the recent acquisition of significant quantities of computed radiography CR equipment for mammography, Mexican regulations do not specify the performance requirements for digital systems such as those of CR type. The design of a quality control program QCP specific for CR mammography systems was thus considered relevant. International protocols were taken as reference to define tests, procedures and acceptance criteria. The designed QCP was applied in three CR mammography facilities. Important deficiencies in spatial resolution, noise, image receptor homogeneity, artifacts and breast thickness compensation were detected.
- OSTI ID:
- 22075688
- Journal Information:
- AIP Conference Proceedings, Vol. 1494, Issue 1; Conference: 12. Mexican symposium on medical physics, Oaxaca (Mexico), 16-19 Mar 2012; Other Information: (c) 2012 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-243X
- Country of Publication:
- United States
- Language:
- English
Similar Records
Radiation protection program for early detection of breast cancer in a mammography facility
WE-A-12A-01: Medical Physics 1.0 to 2.0, Session 2: Radiography, Mammography and Fluoroscopy
SU-E-I-04: A Mammography Phantom to Measure Mean Glandular Dose and Image Quality
Journal Article
·
Fri Nov 07 00:00:00 EST 2014
· AIP Conference Proceedings
·
OSTI ID:22075688
WE-A-12A-01: Medical Physics 1.0 to 2.0, Session 2: Radiography, Mammography and Fluoroscopy
Journal Article
·
Sun Jun 15 00:00:00 EDT 2014
· Medical Physics
·
OSTI ID:22075688
SU-E-I-04: A Mammography Phantom to Measure Mean Glandular Dose and Image Quality
Journal Article
·
Sun Jun 01 00:00:00 EDT 2014
· Medical Physics
·
OSTI ID:22075688