Use of a wire scanner for monitoring residual gas ionization in Soreq Applied Research Accelerator Facility 20 keV/u proton/deuteron low energy beam transport beam line
- SARAF, Soreq Nuclear Research Center, Yavne 81800 (Israel)
The ion source end of the Soreq Applied Research Accelerator Facility accelerator consists of a proton/deuteron ECR ion source and a low energy beam transport (LEBT) beam line. An observed reduction of the radio frequency quadrupole transmission with increase of the LEBT current prompted additional study of the LEBT beam properties. Numerous measurements have been made with the LEBT bream profiler wire biased by a variable voltage. Current-voltage characteristics in presence of the proton beam were measured even when the wire was far out of the beam. The current-voltage characteristic in this case strongly resembles an asymmetric diodelike characteristic, which is typical of Langmuir probes monitoring plasma. The measurement of biased wire currents, outside the beam, enables us to estimate the effective charge density in vacuum.
- OSTI ID:
- 22066315
- Journal Information:
- Review of Scientific Instruments, Vol. 83, Issue 2; Conference: ICIS 2011: 14. international conference on ion sources, Giardini-Naxos, Sicily (Italy), 12-16 Sep 2011; Other Information: (c) 2012 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
- Country of Publication:
- United States
- Language:
- English
Similar Records
2Q-LEBT Prototype for the RIA Facility
CANCELLED Microwave Ion Source and Beam Injection for anAccelerator-Driven Neut ron Source