skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Scintillating screens sensitivity and resolution studies for low energy, low intensity beam diagnostics

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.3488123· OSTI ID:22055734
;  [1]; ; ;  [2]
  1. Cockcroft Institute, Warrington WA4 4AD (United Kingdom)
  2. National Institute of Nuclear Physics INFN-LNS, Catania 95125 (Italy)

In order to investigate the limits of scintillating screens for beam profile monitoring in the ultra-low energy, ultra-low intensity regime, CsI:Tl, YAG:Ce, and a Tb glass-based scintillating fiber optic plate (SFOP) were tested. The screens response to 200 and 50 keV proton beams with intensities ranging from a few picoampere down to the subfemtoampere region was examined. In the following paper, the sensitivity and resolution studies are presented in detail for CsI:Tl and the SFOP, the two most sensitive screens. In addition, a possible use of scintillators for ultra-low energy antiproton beam monitoring is discussed.

OSTI ID:
22055734
Journal Information:
Review of Scientific Instruments, Vol. 81, Issue 10; Other Information: (c) 2010 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States
Language:
English