Title: Design and development of a portable gamma radiation monitor

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.3160292· OSTI ID:22053521
;  [1]; ; ; ;  [2]
  1. Health Physics Division, Directorate of Systems and Services, Pakistan Institute of Nuclear Science and Technology (PINSTECH), Nilore, Islamabad 45650 (Pakistan)
  2. Pakistan Institute of Engineering and Applied Sciences (PIEAS), P. O. Nilore, Islamabad 45650 (Pakistan)

A portable gamma radiation monitor has been designed and developed. The monitor can be used effectively in the dose range from 0.07 to 500 mGy/h due to gamma rays of energy greater than 65 keV. The monitor overestimated radiation doses and the uncertainty in the measured dose rate has been found to be {<=}30%. The response of the monitor can be considered isotropic within an acceptable error of {+-}30%. Provision has also been added to use the monitor as an installed radiation monitor. In installed mode, it can be operated from a remote location up to 1 km and the timing history can be stored on a personal computer.

OSTI ID:
22053521
Journal Information:
Review of Scientific Instruments, Vol. 80, Issue 7; Other Information: (c) 2009 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States
Language:
English