skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Ambient Dose Equivalent measured at the Instituto Nacional de Cancerologia Department of Nuclear Medicine

Journal Article · · AIP Conference Proceedings
DOI:https://doi.org/10.1063/1.3531602· OSTI ID:21506975
 [1];  [1];  [2];  [3];  [4]; ;  [2]
  1. Instituto Nacional de Investigaciones Nucleares, AP 18-1027, 11801, DF (Mexico)
  2. Instituto de Fisica, Universidad Nacional Autonoma de Mexico, AP 20-364, 01000, DF (Mexico)
  3. Instituto Nacional de Cancerologia (INCan), Av. San Fernando 22, C.P. 14080 (Mexico)
  4. Instituto de Ciencias Nucleares, Universidad Nacional Autonoma de Mexico, AP 70-543, 04510 DF (Mexico)

Ambient dose equivalent values were determined in several sites at the Instituto Nacional de Cancerologia, Departmento de Medicina Nuclear, using TLD-100 and TLD-900 thermoluminescent dosemeters. Additionally, ambient dose equivalent was measured at a corridor outside the hospitalization room for patients treated with {sup 137}Cs brachytherapy. Dosemeter calibration was performed at the Instituto Nacional de Investigaciones Nucleares, Laboratorio de Metrologia, to known {sup 137}Cs gamma radiation air kerma. Radionuclides considered for this study are {sup 131}I, {sup 18}F, {sup 67}Ga, {sup 99m}Tc, {sup 111}In, {sup 201}Tl and {sup 137}Cs, with main gamma energies between 93 and 662 keV. Dosemeters were placed during a five month period in the nuclear medicine rooms (containing gamma-cameras), injection corridor, patient waiting areas, PET/CT study room, hot lab, waste storage room and corridors next to the hospitalization rooms for patients treated with {sup 131}I and {sup 137}Cs. High dose values were found at the waste storage room, outside corridor of {sup 137}Cs brachytherapy patients and PET/CT area. Ambient dose equivalent rate obtained for the {sup 137}Cs brachytherapy corridor is equal to (18.51{+-}0.02)x10{sup -3} mSv/h. Sites with minimum doses are the gamma camera rooms, having ambient dose equivalent rates equal to (0.05{+-}0.03)x10{sup -3} mSv/h. Recommendations have been given to the Department authorities so that further actions are taken to reduce doses at high dose sites in order to comply with the ALARA principle (as low as reasonably achievable).

OSTI ID:
21506975
Journal Information:
AIP Conference Proceedings, Vol. 1310, Issue 1; Conference: 11. Mexican symposium on medical physics, Mexico City (Mexico), 19-22 Mar 2010; Other Information: DOI: 10.1063/1.3531602; (c) 2010 American Institute of Physics; ISSN 0094-243X
Country of Publication:
United States
Language:
English

Similar Records

Dose received by occupationally exposed workers at a nuclear medicine department
Journal Article · Tue Oct 23 00:00:00 EDT 2012 · AIP Conference Proceedings · OSTI ID:21506975

Tenth ORNL Personnel Dosimetry Intercomparison Study
Technical Report · Fri Mar 01 00:00:00 EST 1985 · OSTI ID:21506975

'In Vivo' Dosimetry in High Dose Rate Brachytherapy for Cervical Cancer Treatments
Journal Article · Mon Aug 11 00:00:00 EDT 2008 · AIP Conference Proceedings · OSTI ID:21506975