Retrieval of the Protection Filter with a Soft Coaxial Catheter After Carotid Stenting
Journal Article
·
· Cardiovascular and Interventional Radiology
- University of Munich (LMU), Department of Neuroradiology, Klinikum Grosshadern (Germany)
The purpose of this study was to shorten and simplify the procedure of filter retrieval after carotid stenting with cerebral protecting devices. We propose a retrieval method with a soft-tip guiding catheter, which eliminates the step of balloon exchange and introduction of a retriever catheter for the protection filter. The technique was successfully used in our last 60 consecutive patients. One transient deficit occurred. In conclusion, retrieval of the filter with a soft coaxial catheter after carotid stenting is a feasible and safe method and may simplify cerebrovascular protection during stenting.
- OSTI ID:
- 21450312
- Journal Information:
- Cardiovascular and Interventional Radiology, Vol. 32, Issue 1; Conference: BSIR 2008: Annual meeting of the Cardiovascular and Interventional Radiology Society of Europe, Manchester (United Kingdom), 5-7 Nov 2008; Other Information: DOI: 10.1007/s00270-008-9349-x; Copyright (c) 2009 Springer Science+Business Media, LLC; ISSN 0174-1551
- Country of Publication:
- United States
- Language:
- English
Similar Records
Clinical Results of Cerebral Protection with a Filter Device During Stent Implantation of the Carotid Artery
Incidence of New Ischaemic Brain Lesions After Carotid Artery Stenting with the Micromesh Roadsaver Carotid Artery Stent: A Prospective Single-Centre Study
Multiple Coaxial Catheter System for Reliable Access in Interventional Stroke Therapy
Journal Article
·
Sun Jul 15 00:00:00 EDT 2001
· Cardiovascular and Interventional Radiology
·
OSTI ID:21450312
+4 more
Incidence of New Ischaemic Brain Lesions After Carotid Artery Stenting with the Micromesh Roadsaver Carotid Artery Stent: A Prospective Single-Centre Study
Journal Article
·
Tue Nov 15 00:00:00 EST 2016
· Cardiovascular and Interventional Radiology
·
OSTI ID:21450312
+1 more
Multiple Coaxial Catheter System for Reliable Access in Interventional Stroke Therapy
Journal Article
·
Wed Dec 15 00:00:00 EST 2010
· Cardiovascular and Interventional Radiology
·
OSTI ID:21450312
+2 more