skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Endograft Collapse After Endovascular Treatment for Thoracic Aortic Disease

Journal Article · · Cardiovascular and Interventional Radiology
;  [1];  [2];  [3]
  1. Klinikum Wetzlar (Germany)
  2. Klinikum Fulda, Klinik fuer Gefaesschirurgie (Germany)
  3. Klinikum Fulda, Diagnostische und Interventionelle Radiologie (Germany)

Endovascular treatment is an established therapy for thoracic aortic disease. Collapse of the endograft is a potentially fatal complication. We reviewed 16 patients with a thoracic endograft between 2001 and 2006. Medical records of the treated patients were studied. Data collected include age, gender, diagnosis, indication for endoluminal treatment, type of endograft, and time of follow up. All patients (n = 16; mean age, 61 years; range, 21-82 years) underwent computed tomography (CT) for location of the lesion and planning of the intervention. Time of follow-up with CT scan ranged from 1 to 61 months. Indications for endovascular treatment were degenerative aneurysm (n = 7; 44%), aortic dissection (n = 2; 12%), perforated aortic ulcer (n = 4; 25%), and traumatic aortic injury (n = 3; 19%). Three patients suffered from a collapse of the endograft (one patient distal, two patients proximal) between 3 and 8 days after endovascular treatment. These patients were younger (mean age, 37 {+-} 25 years vs. 67 {+-} 16 years; P < 0.05) and showed more oversizing (proximal, 36 {+-} 19.8% vs. 29 {+-} 20.7% [P > 0.05]; distal, 45 {+-} 23.5% vs. 38 {+-} 21.7% [P > 0.05]). Proximal collapse was corrected by placing a bare stent. In conclusion, risk factors for stent-graft collapse are a small lumen of the aorta and a small radius of the aortic arch curvature (young patients), as well as oversizing, which is an important risk factor and is described for different types of endografts and protheses (Gore TAG and Cook Zenith). Dilatation of the collapsed stent-graft is not sufficient. Following therapy implantation of a second stent or surgery is necessary in patients with a proximal endograft collapse. Distal endograft collapse can possibly be treated conservatively under close follow-up.

OSTI ID:
21428980
Journal Information:
Cardiovascular and Interventional Radiology, Vol. 33, Issue 3; Other Information: DOI: 10.1007/s00270-009-9773-6; Copyright (c) 2010 Springer Science+Business Media, LLC and the Cardiovascular and Interventional Radiological Society of Europe (CIRSE); ISSN 0174-1551
Country of Publication:
United States
Language:
English

Similar Records

Long-Term Follow-Up After Endovascular Treatment of Acute Aortic Emergencies
Journal Article · Tue Jan 15 00:00:00 EST 2008 · Cardiovascular and Interventional Radiology · OSTI ID:21428980

CT Imaging Findings and Their Relevance to the Clinical Outcomes After Stent Graft Repair of Penetrating Aortic Ulcers: Six-year, Single-center Experience
Journal Article · Sat Dec 15 00:00:00 EST 2012 · Cardiovascular and Interventional Radiology · OSTI ID:21428980

Endovascular Treatment of Descending Thoracic Aortic Aneurysms with the EndoFit Stent-Graft
Journal Article · Sun Apr 15 00:00:00 EDT 2007 · Cardiovascular and Interventional Radiology · OSTI ID:21428980