skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Pilot Quality Control Program for Audit RT External Beams at Mexican Hospitals

Journal Article · · AIP Conference Proceedings
DOI:https://doi.org/10.1063/1.2979268· OSTI ID:21149270
;  [1]
  1. Secondary Standard Dosimetry Laboratory SSDL, Ionizing Radiation Metrology Department, ININ Carretera Federal Mexico Toluca S/N, La Marquesa, Ocoyoacac, Estado de Mexico 52 750 (Mexico)

A pilot quality control program for audit 18 radiotherapy RT external beams at 13 Mexican hospitals is described--for eleven {sup 60}Co beams and seven photon beams of 6, 10 and 15 MV from accelerators. This program contains five parts: a) Preparation of the TLD-100 powder: washing, drying and annealing (one hour 400 deg. C plus 24 hrs 80 deg. C). b) Sending two IAEA type capsules to the hospitals for irradiation at the hospital to a nominal D{sub W} = 2 Gy{center_dot}c) Preparation at the SSDL of ten calibration curves CC in the range of 0.5 Gy to 6 Gy in terms of absorbed dose to water D{sub W} for {sup 60}Co with traceability to primary laboratory NRC (Canada), according to a window irradiation: 26/10/2007-7/12/2007. d) Reading all capsules that match their hospital time irradiation and the SSDL window irradiation. f) Evaluation of the Dw imparted by the hospitals.

OSTI ID:
21149270
Journal Information:
AIP Conference Proceedings, Vol. 1032, Issue 1; Conference: 10. Mexican symposium on medical physics, Mexico City (Mexico), 17-19 Mar 2008; Other Information: DOI: 10.1063/1.2979268; (c) 2008 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-243X
Country of Publication:
United States
Language:
English