Relative Efficiency Of TLD-100 Exposed to X-Rays
Journal Article
·
· AIP Conference Proceedings
- Instituto de Ciencias Nucleares, Universidad Nacional Autonoma de Mexico, A.P. 70-543, Mexico 04510 DF (Mexico)
- Instituto Nacional de Investigaciones Nucleares, A.P. 18-1027, Mexico 11801 DF (Mexico)
- Instituto de Fisica, Universidad Nacional Autonoma de Mexico, A.P. 20-364, Mexico 01000 DF (Mexico)
The relative efficiency of TLD-100 (LiF:Mg,Ti) exposed to X-rays in the range from 30 to 250 kV, with respect to {sup 60}CO gamma rays has been measured. Glow curves were deconvoluted into peaks 3 to 9. All efficiencies were greater than 1 having a maximum of 1.38 for the dosimetric region (peaks 3+4+5) and peaks 4, 5 and 9, and 2.2 for peaks 6a, 6b, 7 and 8 at the effective energy of 24 keV.
- OSTI ID:
- 21149254
- Journal Information:
- AIP Conference Proceedings, Vol. 1032, Issue 1; Conference: 10. Mexican symposium on medical physics, Mexico City (Mexico), 17-19 Mar 2008; Other Information: DOI: 10.1063/1.2979251; (c) 2008 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-243X
- Country of Publication:
- United States
- Language:
- English
Similar Records
Thermoluminescent response of TLD-100 exposed to low energy protons
Analysis of read-out heating rate effects on the glow peaks of TLD-100 using WinGCF software
Response of LiF:Mg,Ti thermoluminescent dosimeters at photon energies relevant to the dosimetry of brachytherapy (<1 MeV)
Journal Article
·
Fri Aug 28 00:00:00 EDT 1998
· AIP Conference Proceedings
·
OSTI ID:21149254
+4 more
Analysis of read-out heating rate effects on the glow peaks of TLD-100 using WinGCF software
Journal Article
·
Fri Jan 22 00:00:00 EST 2016
· AIP Conference Proceedings
·
OSTI ID:21149254
Response of LiF:Mg,Ti thermoluminescent dosimeters at photon energies relevant to the dosimetry of brachytherapy (<1 MeV)
Journal Article
·
Sat Oct 15 00:00:00 EDT 2011
· Medical Physics
·
OSTI ID:21149254
+1 more